Gas detection within energy storage compartments using cavity-enhanced Raman spectroscopy with a self-adaptive baseline removal and signal temperature correction.
Zhiwei Shen, Pinyi Wang, Miaolin Wang, Longtao He, Weigen Chen
State Key Laboratory of Power Transmission Equipment & System Security and New Technology, Chongqing University, Chongqing, 400044, China; National Innovation Center for Industry Education Integration of Energy Storage Technology, Chongqing, 400044, China. · State Key Laboratory of Power Transmission Equipment & System Security and New Technology, Chongqing University, Chongqing, 400044, China; National Innovation Center for Industry Education Integration of Energy Storage Technology, Chongqing, 400044, China. Electronic address: wang.pinyi@qq.com.
Key points
From the paper's conclusion- The automated, highly sensitive analysis of multiple gases with temperature correction via Raman spectroscopy demonstrates the practicality of CERS for gas detection in energy storage compartments.
Abstract
Accurate analysis of the gases (mainly including CH4, H2, CO, and CO2) within electrochemical energy storage compartments is of paramount importance for ensuring the safe operation of batteries. For this purpose, we propose a cavity-enhanced Raman spectroscopy gas sensing method with a self-adaptive baseline removal and signal temperature correction. Based on an open-source low-pass filter-based baseline fitting method (BEADS), we develop a key parameter (cut-off frequency) selection algorithm without manual intervention to obtain Raman information accurately and automatically. The effectiveness of this method is validated through the analysis of Raman intensity and shifts in an artificial Raman spectrum. Besides, the multi-pass cavity enhanced Raman spectroscopy is employed to investigate the temperature-dependent variations in the information of Raman signals (Raman intensity, shift, and linewidth) of typical gases within the battery compartment. Based on the variation of intensity, a signal intensity calibration method for different temperatures is derived. The automated, highly sensitive analysis of multiple gases with temperature correction via Raman spectroscopy demonstrates the practicality of CERS for gas detection in energy storage compartments.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Tessellated-Pattern Dense-Pattern Multi-Pass Cavity Enhanced Raman Spectroscopy for In Situ Analysis of Gases Inside Lithium-Ion BatteriesAnalytical Chemistry · 2026
- Herriott-type cavity-enhanced Raman spectroscopy for simultaneous analysis of ten natural gas components.Spectrochimica acta. Part A, Molecular and biomolecular spectroscopy · 2026
- Detecting Decomposition Gases from Gas-Insulated Switchgear with Cavity-Enhanced Polarization-Differential Raman SpectroscopyAnalytical Chemistry · 2026
- Rapid real-time monitoring of dissolved gases in seawater by cavity-enhanced Raman spectroscopy system with sub-ppm sensitivity.Talanta · 2026
- Quantitative Analysis of Multicomponent Converter Gas Using Raman Spectroscopy with Light Intensity Correction.Analytical chemistry · 2026
- Modified lithium-ion cell design enabling operando gas sensing using optical fiber photothermal spectroscopyFrontiers in Energy Research · 2026