Talanta

Online extraction and preconcentration system employing a mini-column packed with 4-(2-thiazolylazo)-resorcinol-coated CoFe2O4 nanoparticles for the determination of lead in water samples by F AAS.

José J C L de Souza Júnior, Bruna R S Santos, Geovani C Brandão, Sérgio L C Ferreira, Walter N L Dos Santos, Leonardo B Guimarães

Universidade do Estado da Bahia, Departamento de Ciências Exatas e da Terra, Cabula, 41150-000, Salvador, BA, Brazil. · Universidade Federal da Bahia, Instituto de Química, Campus de Ondina, 40170-290, Salvador, BA, Brazil.

Key points

From the paper's conclusion
  • In this study, an online solid-phase extraction system was developed that employs a mini-column packed with 4-(2-thiazolylazo)-resorcinol (TAR)-coated CoFe2O4 nanoparticles for the preconcentration and determination of lead in water samples by flame atomic absorption spectrometry.

Abstract

In this study, an online solid-phase extraction system was developed that employs a mini-column packed with 4-(2-thiazolylazo)-resorcinol (TAR)-coated CoFe2O4 nanoparticles for the preconcentration and determination of lead in water samples by flame atomic absorption spectrometry. The synthesis of CoFe2O4 nanoparticles was accomplished through a coprecipitation method in basic media, followed by a coating process with the TAR complexing reagent. The characterization of both uncoated and coated nanoparticles was performed by means of analysis of X-ray diffraction (XRD), Fourier-transform infrared (FT-IR) spectroscopy, and transmission electron microscopy (TEM). Multivariate methodology was employed to optimize the flow injection analysis (FIA) system. A two-level full factorial design was employed to evaluate the factors, namely eluent concentration, sampling flow rate, and pH of the sample or standard solution. The Box-Behnken design was utilized to ascertain the optimal conditions for the online system. The eluent concentration was determined to be 0.10 mol L-1, the sampling flow rate was set to 2.5 mL min-1, and the pH was adjusted to 5.0. For a sample volume of 25.0 mL and under the established optimum conditions, the online system has been shown to achieve a limit of detection and quantification of 0.9 and 2.9 μg L-1, respectively, with an enrichment factor of 71 and a frequency of 5.7 h-1 for the determination of lead in water. The precision of the proposed method, evaluated as relative standard deviation (RSD%), yielded values of 7.3% and 6.9% for standard solutions containing 10.0 and 50.0 μg L-1 of Pb, respectively, while intraday and interday RSD values remained below 5.3% and 8.1%. The accuracy of the method was confirmed through analysis of the certified reference materials European Ultrapure Laboratory water (EUL, CRM) and European Primary Laboratory water (EPL, CRM), with an agreement ranging from 90% to 102%. The proposed flow injection analysis line system was subsequently employed for the analysis of mineral water and tap water. To evaluate the method's accuracy, an additional recovery test was also performed. The recovery values obtained ranged from 94 to 103%.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers