Development of a reliable microwave plasma-atomic emission spectrometry (MP-AES) method for the determination of total arsenic in groundwater using l-cysteine as a reducing agent.
Shamim Ahmed, Warren Bruckard, Kim Dowling, Samina Ahmed, Yalong Ma, Miao Chen
School of Science, RMIT University, City Campus, VIC, 3001, Australia; CSIRO Mineral Resources, Clayton, Vic 3168, Australia; Bangladesh Council of Scientific and Industrial Research (BCSIR), Bangladesh. · School of Science, RMIT University, City Campus, VIC, 3001, Australia.
Key points
From the paper's conclusion- The results were in excellent agreement with those obtained by conventional inductively coupled plasma mass spectrometry (ICP-MS), demonstrating the method's suitability for accurate determination of total arsenic in groundwater.
Abstract
Accurate determination of arsenic in groundwater is critically important due to the widespread occurrence of arsenic contamination worldwide. This study aimed to develop a reliable and affordable analytical method for the determination of total arsenic in groundwater. The method uses hydride generation microwave plasma atomic emission spectrometry (HG-MP-AES) coupled with a multimode sample introduction system (MSIS), with l-cysteine as the reducing agent. The results demonstrated that accurate determination of total arsenic, comprising both As(III) and As(V), requires the complete reduction of As(V) to As(III) prior to analysis. This was achieved by treating samples with 1% l-cysteine in a 1% hydrochloric acid (HCl) solution for 3 h. The effectiveness of the reduction process was verified by high-performance liquid chromatography coupled with inductively coupled plasma mass spectrometry (HPLC-ICP-MS) analysis. The developed method provides reliable and accurate determination of total arsenic in groundwater and offers a practical analytical approach for the routine monitoring of arsenic-contaminated water resources. The method achieved a limit of detection (LOD) of 0.49 ppb and a limit of quantification (LOQ) of 1.47 ppb. Spike recovery values for groundwater samples ranged from 98% to 107%, while the relative standard deviation (RSD) was below 5%, demonstrating excellent accuracy and precision. A total of 40 groundwater samples collected from two different regions of Bangladesh were analysed using the developed method. The results were in excellent agreement with those obtained by conventional inductively coupled plasma mass spectrometry (ICP-MS), demonstrating the method's suitability for accurate determination of total arsenic in groundwater.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Arsenic in aquatic and sedimentary environments: Speciation, mobility, analytical advances, and implications for risk-based monitoringEcological Engineering & Environmental Technology · 2026
- Sensitive determination of arsenic in water using fluorescent CdS nanostructures.The Analyst · 2026
- Analytical techniques and source apportionment for heavy metal(loid)s in groundwater: A comprehensive reviewTalanta Open · 2025
- Field-Deployable and Ultrasensitive Surface-Enhanced Raman Spectroscopy for Waterborne Arsenic Monitoring Based on Headspace Formation of Arsenic-Silver Nanocomposites.ACS sensors · 2026
- Emergence of Electroanalytical Techniques for Arsenic Detection in Water: Recent Trends and Future PerspectivesChemElectroChem · 2025
- Electroanalytical overview: recent advances in the sensing of arsenic using screen-printed electrochemical platforms.Environmental science. Nano · 2026