Outcomes after rapid introduction of new surgical staplers: "Your gut says one thing, the data say another".
Isabelle M C Tan, Julie Hong, Daniel A Metzger, Shahenda Khedr, Julianna Brouwer, Ying Li, Steven Y Chao
Department of Surgery, Weill Cornell Medicine, NewYork-Presbyterian Hospital, New York, NY; Stritch School of Medicine, Loyola University of Chicago, Maywood, IL. · Department of Surgery, NewYork-Presbyterian Queens, Flushing, NY.
Key points
From the paper's conclusion- Adoption of new surgical staplers was not associated with changes in perioperative outcomes. Further analysis of rollout strategies, such as the duration of in-service teaching and qualitative feedback on implementation, could inform best practices for safely adopting surgical innovation. While anecdotal experiences may bias surgeons against change, these results support the evidence-based adoption of new surgical devices.
Abstract
BACKGROUND: Hospitals frequently change surgical equipment for cost, safety, and efficiency. However, the impact of adopting new technology on postoperative outcomes is not well characterized. This study examined whether transitioning to a new brand of surgical staplers was associated with changes in surgical outcomes. METHODS: A single-institution retrospective cohort study of 1,211 surgeries performed between January 3, 2022 and November 22, 2024 was conducted using American College of Surgeons National Surgical Quality Improvement Program data. On November 13, 2023, Medtronic staplers replaced all Ethicon staplers (n = 867 pretransition; n = 344 posttransition). Operations routinely utilizing staplers affected by the switch were included. Outcomes measured included postoperative complications, blood transfusions, surgical site infection, return to the operating room, 30-day readmission, discharge to home, and hospital length of stay. Multivariable analyses were performed, adjusting for patient demographics and case characteristics. Analyses of appendectomy and colectomy subgroups were also performed. RESULTS: Adjusted analysis demonstrated no change in postoperative complications (adjusted odds ratio, 0.606; P = .052), blood transfusion (adjusted odds ratio, 1.051; P = .917), surgical site infection (adjusted odds ratio, 1.108; P = .741), return to operating room (adjusted odds ratio, 0.584; P = .365), 30-day readmission (adjusted odds ratio, 0.690; P = .326), discharge to home (adjusted incidence rate ratio, 0.975; P = .682), hospital length of stay (adjusted incidence rate ratio, 0.986; P = .774), postoperative length of stay (adjusted incidence rate ratio, 0.975; P = .682), and surgery duration (adjusted incidence rate ratio, 0.989; P = 0.603). Subgroup analyses showed no differences in outcomes. CONCLUSION: Adoption of new surgical staplers was not associated with changes in perioperative outcomes. Further analysis of rollout strategies, such as the duration of in-service teaching and qualitative feedback on implementation, could inform best practices for safely adopting surgical innovation. While anecdotal experiences may bias surgeons against change, these results support the evidence-based adoption of new surgical devices. LEVEL OF EVIDENCE: III, Retrospective cohort study.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Robotic vs. bedside staplers in robotic lung resection: a real-world cohort study.Journal of robotic surgery · 2026
- Access without excess: a comparison of wound infections with disposable versus hybrid trocars in minimally invasive inguinal hernia repairs.Surgical endoscopy · 2026
- Robotic vs. Laparoscopic Stapling in Robotic Total Mesorectal Excision: A Prospective Cohort Study.Cureus · 2026
- Hautmann versus modified Studer orthotopic neobladder after radical cystectomy.BJU international · 2026
- The Hospital Matters: Association Between Hospital Teaching Status and Patient Outcomes in Emergency General Surgery.The Journal of surgical research · 2026
- Preventive Carbapenem Stewardship and Clinical Outcomes in a Low-Resistance Setting: A Multicenter Quasi-Experimental Study.International journal of antimicrobial agents · 2026