Rod Geometry and Material Characteristics as Risk Factors for Rod Breakage After Adult Spinal Deformity Surgery: A Multicenter Cohort Study with Five-Year Follow-up.
Alice Baroncini, Daniel Larrieu, Louis Boissiere, Cecile Roscop, Ferran Pellisé, Ahmet Alanay, Frank Kleinstueck, Javier Pizones,
Department of Spine Surgery I, Casa di Cura Humanitas San Pio X, Milan, Italy. Electronic address: Alice.baroncini@gmail.com. · ELSAN Polyclinique Jean Villar-French Spine Study Group (FSSG), Bruges, France.
Key points
From the paper's conclusion- Conclusions Specific rod geometry characteristics-particularly early lower lumbar curvature and limited postoperative curvature increase-were independently associated with rod breakage after ASD surgery.
Abstract
Background Context Rod breakage is a frequent and clinically important late complication after adult spine deformity (ASD) surgery, yet most research has focused on patient and alignment factors rather than implant geometry. Purpose This study aimed to investigate the rod-related characteristics independently associated with rod breakage following ASD surgery, with particular focus on quantitative measures of rod material, rod shape and its postoperative evolution. Design Retrospective analysis of prospectively collected, six-center data. Patient Sample ASD patients, who underwent long posterior fusion to the pelvis with available serial full-length lateral radiographs at 6 weeks, 6 months, 1 year, and 2 years were included; patients with multiple-rod constructs were excluded. Minimum two-year follow-up. Outcome Measures Rod breakage (defined radiographically), time to rod breakage, and discrimination performance of a multivariable predictive model based on patient and rod characteristics. Methods Rod curvature was quantified at L1-S1 and L4-S1 and the maximum positive increase in L1-S1 curvature between consecutive postoperative timepoints was derived. Time-to-breakage was analyzed using Kaplan-Meier and Cox proportional hazards models, with data-driven thresholds defining high-risk geometry. Model discrimination was assessed using the concordance index. Results Eighty-three patients (mean follow-up 60.0 ± 23.0 months) were analyzed; 30 developed rod breakage at a mean 41.7 ± 22.2 months. L4-S1 rod curvature ≥33° at 6 weeks was associated with higher risk of breakage (HR 2.25; 95% CI 1.06-4.77; p = 0.036), as was a maximum L1-S1 curvature increase≤4° (HR 3.33; 95% CI 1.25-8.84; p = 0.016). Titanium rods had lower risk compared with cobalt-chromium (HR for titanium 0.50; p = 0.02), and diameter >5.5 mm was associated with increased risk (HR 1.90; p = 0.02). A multivariable patient-level model incorporating rod geometry showed improved discrimination (C-index 0.74). Conclusions Specific rod geometry characteristics-particularly early lower lumbar curvature and limited postoperative curvature increase-were independently associated with rod breakage after ASD surgery. Integrating rod shape into risk models may support implant selection, surgical planning, and postoperative monitoring. Level of Evidence Level III, retrospective cohort study.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Combined Anterior-Posterior vs. Posterior-Only Fusion in Adult Spinal Deformity Radiographic, Clinical, and Mechanical Outcomes With Longitudinal Mixed-Effects Analysis.Global spine journal · 2026
- Patterns of later-stage growing rod fracture and their relationship with thoracic hyperkyphosis in early onset scoliosis.Journal of orthopaedic surgery and research · 2026
- Long-term impact of anterior longitudinal ligament rupture on rod fracture and fusion status after anterior and posterior repair with lateral lumbar interbody fusion for adult spinal deformity.Spine deformity · 2026
- Distinct Mechanisms of Proximal Junctional Kyphosis and Their Clinical ImplicationsSpine · 2026
- Tether Breakage after Vertebral Body Tethering: A Systematic Review and Meta-Analysis of Time-dependent Risk and Clinical Consequences.Global spine journal · 2026
- Postoperative coronal imbalance predicts late mechanical failure and revision surgery after adult spinal deformity correction with pelvic fixation: a 5-year follow-up study in elderly patients.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2026