Proactively identifying real world risks during healthcare technology implementation: A case study using FMEA and usability studies.
Archie Lodge, Adam Mackridge, Christine Bond, Peter Butcher, Rebecca Payne
Nuffield Department of Primary Care Health Sciences, University of Oxford, United Kingdom. · North Wales Medical School, Bangor University, United Kingdom; Betsi Cadwaladr University Health Board, United Kingdom.
Key points
From the paper's conclusion- The combined use of an adapted FMEA and contextual usability study enabled proactive identification and mitigation of system reliability and patient safety risks prior to real-world deployment.
Abstract
BACKGROUND: Implementing new technologies in healthcare is complex, failure-prone and can introduce unintended risks to patient safety. Structured risk management methods such as Failure Modes and Effects Analysis (FMEA) and usability testing are widely used in safety-critical industries prior to technology implementation, but infrequently in healthcare. OBJECTIVES: To use FMEA and usability testing techniques to proactively identify and mitigate patient safety and implementation risks prior to novel deployment of an existing healthcare technology: a modification of a machine to issue medication directly to a patient, following an out-of-hours teleconsultation. METHODS: A prospective case study was undertaken during the pre-implementation phase of a remotely controlled medication-issuing machine designed to improve out-of-hours access to urgent medicines in a rural area of North Wales. The machine was based on an existing medication-issuing system, modified for a new urgent-care pathway. A pragmatically adapted FMEA was conducted across three stages of system use (machine loading, remote clinician authorisation, and patient collection) using structured multidisciplinary workshops. This was followed by two rounds of contextual usability studies involving direct observation and structured interviews with members of the public (pseudo-patients) interacting with the machine in real-world conditions. The assessment focused on medication access after a clinical decision had been made and did not evaluate upstream triage or prescribing decisions. Risks and usability issues were identified and mitigations adopted. RESULTS: The FMEA identified multiple previously unrecognised risks to patient safety and system reliability, including the potential for medication loading errors, failures in PIN code delivery, and lack of contingency for technical faults. Usability studies involving 45 participants identified additional usability issues not detected by FMEA, disproportionately affecting older and disabled users. CONCLUSIONS: The combined use of an adapted FMEA and contextual usability study enabled proactive identification and mitigation of system reliability and patient safety risks prior to real-world deployment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- An Evaluation of Clinician and Informatician Perspectives on Medication Harm Risk Prediction Tools for Use in a Tertiary Hospital Setting.Journal of evaluation in clinical practice · 2026
- Prospective Failure Mode and Effects Analysis to Support Implementation Readiness for a New Hyperbaric Oxygen Therapy Service.Joint Commission journal on quality and patient safety · 2026
- Testing Failure Mode and Effect Analysis: A Descriptive Empirical Study in Paediatric and Adolescent Hospital CareJournal of Evaluation in Clinical Practice · 2026
- APPLICATION OF THE FAILURE MODE AND EFFECTS ANALYSIS METHOD IN HEALTHCARE: A CASE STUDY OF A UNIVERSITY MEDICAL CENTER IN KAZAKHSTAN.Georgian medical news · 2026
- Red Teaming a Patient-Facing Clinical AI Platform: 2-Phase Prospective Study.JMIR AI · 2026
- Clinical AI in Canadian Community Hospitals: A Framework for Implementation Readiness.Healthcare management forum · 2026