Characteristics of pediatric central venous catheter malfunction and subsequent catheter-related deep vein thrombosis.
Natasha S Bala, Zaineb Boulil, Emily Krava, Char Witmer, Sarah H O'Brien, Lingyun Ji, Guy Young, Julie Jaffray
Center for Cancer and Blood Disorders, Department of Pediatrics, Phoenix Children's Hospital, Phoenix, Arizona, USA. · Research Administration Department, Rady Children's Hospital, San Diego, California, USA.
Key points
From the paper's conclusion- CVC malfunction is common and associated with several modifiable catheter-related factors. Six percent of malfunctioning CVCs develop CR-DVT, and repeated t-PA use and persistent CVC dysfunction may warrant further evaluation for underlying thrombosis.
Abstract
BACKGROUND: Central venous catheters (CVCs) are increasingly placed in children; however, malfunction and deep vein thrombosis (DVT) are reasons for frequent removal. OBJECTIVES: The objective of this study was to evaluate variables associated with CVC malfunction and, subsequently, the development of CVC-related DVT (CR-DVT). METHODS: This subanalysis of the multicenter, prospective Clot Incidence Rates in Central Lines study enrolled children aged 6 months to <18 years with newly placed peripherally inserted central catheters or tunneled lines from 2013 to 2018. Time from CVC placement to first malfunction was analyzed in the full cohort using univariable and multivariable Weibull regression. Among CVCs that malfunctioned, time from first malfunction to CR-DVT was analyzed to assess tissue plasminogen activator (t-PA) use and subsequent thrombosis. RESULTS: Malfunction occurred in 380/1951 CVCs (19.5%). In multivariable analysis, prior CVC use (hazard ratio [HR], 1.51; 95% CI, 1.19-1.90; P = .001), multilumen catheters (HR, 2.89; 95% CI, 2.30-3.65; P < .001), and >2 insertion attempts (HR, 2.21; 95% CI, 1.42-3.45; P < .001) were associated with a higher risk of malfunction, whereas catheter size >4 French and interventional radiology placement were associated with a lower risk. Among malfunctioning CVCs, 22/380 (5.8%) developed CR-DVT, with a median time to CR-DVT of 6 days after malfunction. Increasing t-PA use showed a nonsignificant trend toward higher subsequent CR-DVT. CONCLUSION: CVC malfunction is common and associated with several modifiable catheter-related factors. Six percent of malfunctioning CVCs develop CR-DVT, and repeated t-PA use and persistent CVC dysfunction may warrant further evaluation for underlying thrombosis.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clinical performance and microstructural assessment of hemodialysis catheters for children on maintenance dialysis.Pediatric nephrology (Berlin, Germany) · 2026
- Factors Associated with Venous Catheter Related Complications and Their Implications for Improving Patient Care: An Observational StudyNursing Reports · 2026
- Complications of Central Venous Catheters in Egyptian Children and Adolescents With Malignancy: A Prospective Single-Center Observational Study.Journal of pediatric hematology/oncology · 2026
- Complications and predictors of catheter failure in pediatric long peripheral and midline catheters: a prospective observational study.Journal of infection prevention · 2026
- Epidemiological Characteristics of Venous Thromboembolism and Associated Indicators for Catheter-Related Thrombosis Subtypes.Journal of visualized experiments : JoVE · 2026
- Peripherally inserted central catheter-related complications: A national prospective multicenter study of incidence, risk factors and outcomes in China.The journal of vascular access · 2026