Respiratory medicine

Development, content validation and reliability testing of a measure of inhaler technique for all devices: The Portsmouth Inhaler Technique (PIT) Score.

Ruth De Vos, Mitch Lomax, Heather Mackenzie, Jayne Longstaff, Ina Kostakis, Milan Chauhan, Robin Clark, Daniel Neville,

University of Portsmouth; Portsmouth Hospitals University NHS Trust. · University of Portsmouth.

Guideline

Key points

From the paper's conclusion
  • To our knowledge, the PIT Score is the first reliable clinician-observed assessment tool applying one set of scored steps to all inhaler types. Its ten scored steps give a practical, standardised record of technique across devices. Prospective evaluation in routine practice, in children and against clinical outcomes is needed.

Abstract

BACKGROUND: Inhaler technique errors are common and associated with poorer disease control. Clinicians assess technique against device-specific checklists that differ in content, limiting comparisons across inhaler types. OBJECTIVE: To develop and evaluate the Portsmouth Inhaler Technique (PIT) Score, a standardised, device-neutral assessment tool for adults and children. METHODS: Inhaler technique steps were extracted from 418 published and NHS checklists and underwent three rounds of Delphi essentiality assessment by 18 experts, followed by relevance (34 experts) and clarity (62 assessors, including inhaler users) evaluation. Discriminant validity was examined by comparing PIT Scores, derived retrospectively for 677 adults in an observational study, with Asthma Control Questionnaire (ACQ6) or COPD Assessment Test (CAT) scores. Twenty-two experts assessed 14 standardised videos representing seven different inhaler devices to evaluate reliability; 20 repeated the assessment. RESULTS: Seventy-two steps were refined over three rounds into a 16-step checklist with ten scored and six unscored steps. All steps met the relevance threshold (I-CVI ≥0.78); average scale-level content validity was 0.95. PIT Scores showed no meaningful correlation with ACQ6 (ρ=0.008; p=0.858) or CAT (ρ=0.077; p=0.267), supporting discriminant validity. Inter-rater reliability was excellent (Gwet's AC1=0.9177; ICC=0.9015), as was intra-rater agreement (AC1>0.92). Median PIT Score completion time was 58 seconds. CONCLUSIONS: To our knowledge, the PIT Score is the first reliable clinician-observed assessment tool applying one set of scored steps to all inhaler types. Its ten scored steps give a practical, standardised record of technique across devices. Prospective evaluation in routine practice, in children and against clinical outcomes is needed.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers