Identifying the Operational Bottleneck to Shock Delivery After Automated External Defibrillator Activation.
Tsuyoshi Nojima, Tetsuya Yumoto, Takafumi Obara, Kohei Tokioka, Takashi Hongo, Yoshinori Kosaki, Kohei Tsukahara, Atsunori Nakao,
Okayama University Faculty of Medicine, Dentistry, and Pharmaceutical Sciences, Department of Emergency, Critical Care, and Disaster Medicine, 2-5-1 Shikatacho, Kita-ku, Okayama 700-8558, Japan. Electronic address: t.nojima1002@gmail.com. · Okayama University Faculty of Medicine, Dentistry, and Pharmaceutical Sciences, Department of Emergency, Critical Care, and Disaster Medicine, 2-5-1 Shikatacho, Kita-ku, Okayama 700-8558, Japan.
Key points
From the paper's conclusion- The interval from AED power-on to electronically detected completion of pad application, encompassing all intervening actions, was the largest measurable post-activation bottleneck. Completion within 50 seconds may serve as an exploratory process target for first recorded shock delivery within 60 seconds among initially shock-advised events.
Abstract
INTRODUCTION: Cardiac arrest registries record automated external defibrillator (AED) use and shock delivery but not post-activation operational timing. We aimed to identify the largest measurable bottleneck and an exploratory pad-application target for shock delivery within 60 seconds of AED power-on. METHODS: We retrospectively analyzed timestamps from remotely monitored Nihon Kohden AEDs between January 2018 and February 2026. After excluding inspection records, we grouped records into activation events and quantified intervals. Shock-timing and receiver operating characteristic (ROC) analyses were restricted to initially shock-advised events with recorded shock delivery. RESULTS: We analyzed 8,115 activation events. Median times were 42 seconds (interquartile range [IQR], 30-56) from power-on to completion of pad application, nine seconds (IQR, 9-11) from pad application to the first analysis, and 52 seconds (IQR, 40-66) from power-on to the first analysis endpoint. The power-on-to-pad-application interval accounted for a median of 81% (IQR, 75-85) of the total power-on-to-first-analysis interval, increasing to 87% (IQR, 85-90) when first analysis occurred more than 60 seconds after power-on. Among 900 initially shock-advised events with recorded shock delivery, the median pad-application-to-shock interval was 15 seconds (IQR 13-17). Exploratory ROC analysis yielded a power-on to pad-application threshold of ≥ 51 seconds in relation to first recorded shock delivery occurring more than 60 seconds after AED power-on. CONCLUSIONS: The interval from AED power-on to electronically detected completion of pad application, encompassing all intervening actions, was the largest measurable post-activation bottleneck. Completion within 50 seconds may serve as an exploratory process target for first recorded shock delivery within 60 seconds among initially shock-advised events.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- AnterioR-Posterior VErsuS Anterior-LaTeral defibrillator pAd position in out of hospital cardiac aRresTmedRxiv · 2026
- Out-of-Hospital Cardiac Arrest: Public-Access Defibrillation and System Approaches to Minimize Avoidable DelayJournal of Clinical Medicine · 2026
- Superior-head automated external defibrillator placement during single-rescuer cardiopulmonary resuscitation: a randomized crossover simulation study and retrospective multicenter cohort.Frontiers in public health · 2026
- The early defibrillation gap: a time-dependent systems model of survival loss in out-of-hospital cardiac arrestEuropean Heart Journal · 2026
- A randomised controlled trial of defibrillation with manual pressure augmentation during out-of-hospital cardiac arrestResuscitation · 2026
- Community First Responder Activation Reduces the Negative Impact of AED Distance on Bystander Defibrillation After Out-of-Hospital Cardiac Arrest.Resuscitation · 2026