Minimal.flow Autocontrol Anaesthesia System (MAAS): A Clinical Model for Low-flow Anaesthesia. A Prospective Paired Study with Sevoflurane and Desflurane.
M de la Matta, S López Bustos, A Domínguez Blanco
Servicio de Anestesiología, Reanimación y Tratamiento del Dolor, Hospital Universitario Virgen del Rocío, Sevilla, Spain. Electronic address: mdlmatta@hotmail.com. · Servicio de Anestesiología, Reanimación y Tratamiento del Dolor, Hospital Universitario Virgen del Rocío, Sevilla, Spain.
Key points
From the paper's conclusion- The MAAS method allows simple and effective maintenance of the FeVA target with both sevoflurane and desflurane while using low fresh gas flow from the start of volatile anaesthetic administration. These findings support its clinical applicability as a practical and efficient tool for low flow anaesthesia.
Abstract
INTRODUCTION: The Minimal flow Autocontrol Anaesthesia System (MAAS) uses the inspiratory to expiratory halogenated anaesthetic (HA) gradient to estimate HA requirements during low flow anaesthesia. Its clinical applicability has not been evaluated in the literature. The aim of this study was to compare the accuracy of the MAAS in maintaining the target alveolar concentration of halogenated anaesthetics (FeHA target) when using sevoflurane vs desflurane. METHODS: This was a prospective, paired, observational study involving 29 patients undergoing robotic surgery under general anaesthesia. The fresh gas flow was maintained at 10% of minute volume throughout the procedure. Sevoflurane and desflurane were administered sequentially. The number of vaporiser adjustments required to maintain a stable FeHA target during maintenance, the time required to reach the FeHA target, the duration of FeHA and BIS target deviations, and adjustments in oxygen delivery were analysed. RESULTS: No differences were observed in the number of vaporiser adjustments, although there was a trend towards a higher number of adjustments with desflurane during the first 15 min. The cumulative duration of FeHA target deviations was less than 2% in both agents. No differences were found in the time required to achieve FeHA target or in oxygen delivery adjustments. CONCLUSIONS: The MAAS method allows simple and effective maintenance of the FeVA target with both sevoflurane and desflurane while using low fresh gas flow from the start of volatile anaesthetic administration. These findings support its clinical applicability as a practical and efficient tool for low flow anaesthesia.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Consumption of isoflurane and recovery time in low-flow anaesthesia versus high-flow anaesthesia.Pakistan journal of medical sciences · 2026
- Recovery Profile Following Isoflurane-based Low-flow Anaesthesia with Nitrous Oxide and without Nitrous Oxide: An Observational Cohort study.The Nigerian postgraduate medical journal · 2026
- Remimazolam versus sevoflurane for general anaesthesia: A meta-analysis with trial sequential analysis of haemodynamic and clinical outcomes.European journal of anaesthesiology · 2026
- Comparative Evaluation of the Vasodilatory Effects of Desflurane, Isoflurane, and Sevoflurane Using Perfusion Index During General Anaesthesia.Cureus · 2026
- Evaluation of the efficacy of sevoflurane as a general anesthetic method for noninvasive examination of gentoo penguins (Pygoscelis papua).The Journal of veterinary medical science · 2026
- More vasopressors were required during anesthesia maintenance with sevoflurane than remimazolam in a high-risk patient: a case report.JA clinical reports · 2026