  • The MAAS method allows simple and effective maintenance of the FeVA target with both sevoflurane and desflurane while using low fresh gas flow from the start of volatile anaesthetic administration. These findings support its clinical applicability as a practical and efficient tool for low flow anaesthesia.

Abstract

INTRODUCTION: The Minimal flow Autocontrol Anaesthesia System (MAAS) uses the inspiratory to expiratory halogenated anaesthetic (HA) gradient to estimate HA requirements during low flow anaesthesia. Its clinical applicability has not been evaluated in the literature. The aim of this study was to compare the accuracy of the MAAS in maintaining the target alveolar concentration of halogenated anaesthetics (FeHA target) when using sevoflurane vs desflurane. METHODS: This was a prospective, paired, observational study involving 29 patients undergoing robotic surgery under general anaesthesia. The fresh gas flow was maintained at 10% of minute volume throughout the procedure. Sevoflurane and desflurane were administered sequentially. The number of vaporiser adjustments required to maintain a stable FeHA target during maintenance, the time required to reach the FeHA target, the duration of FeHA and BIS target deviations, and adjustments in oxygen delivery were analysed. RESULTS: No differences were observed in the number of vaporiser adjustments, although there was a trend towards a higher number of adjustments with desflurane during the first 15 min. The cumulative duration of FeHA target deviations was less than 2% in both agents. No differences were found in the time required to achieve FeHA target or in oxygen delivery adjustments. CONCLUSIONS: The MAAS method allows simple and effective maintenance of the FeVA target with both sevoflurane and desflurane while using low fresh gas flow from the start of volatile anaesthetic administration. These findings support its clinical applicability as a practical and efficient tool for low flow anaesthesia.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers