Uncertainty quantification for detecting out-of-distribution cases in deep learning segmentation of head and neck organs of interest.
Joëlle E van Aalst, Rita Simões, Federica C Maruccio, Daniel C MacRae, Luuk van der Hoek, Johannes A Langendijk, Peter M A van Ooijen, Tomas M Janssen,
University of Groningen, University Medical Center Groningen, Department of Radiation Oncology, Groningen, the Netherlands. Electronic address: j.e.van.aalst@umcg.nl. · The Netherlands Cancer Institute, Department of Radiation Oncology, Amsterdam, the Netherlands.
Key points
From the paper's conclusion- Uncertainty quantification can detect distribution shifts in radiotherapy auto-segmentation. This could support automated QA workflows by identifying in-distribution cases and enabling the application of specialised models.
Abstract
PURPOSE: Quality assurance (QA) of auto-segmentations is necessary for safe clinical deployment. Deep learning models are optimised for data similar to their training set, but reliability drops with out-of-distribution (OOD) inputs. This study evaluates whether uncertainty quantification (UQ) can detect OOD inputs arising from distribution shift in head and neck organ of interest (OOI) auto-segmentation. MATERIALS AND METHODS: Monte Carlo dropout-based uncertainty scores and geometric accuracy of nnU-Net models were evaluated in two experiments. First, as a proof-of-concept, we assessed whether uncertainty scores detect metal artefact (N = 10) from non-artefact cases (N = 10) on a baseline model (mixed training data, N = 620) and a no-metal model (artefact-free training data, N = 620). Second, in a clinically realistic experiment, we split a fine-tuning dataset by uncertainty score to obtain a narrow model (low-uncertainty cases, N = 21) and a broad model (high-uncertainty cases, N = 21). A held-out test set (N = 76) evaluated whether segmentation accuracy differed between uncertainty categories. RESULTS: Both experiments showed that uncertainty scores reflect the learned training distribution. In the supervised experiment, the no-metal model demonstrated greater uncertainty score separation between artefact and non-artefact cases (1.1 × 105 vs 0.6 × 105) than the baseline model (1.1 × 105 vs 0.8 × 105). In the unsupervised experiment, the narrow model performed significantly better on certain than uncertain cases (p = 0.04), while no difference was seen for the broad model (p = 0.34), which did include these cases in training. CONCLUSIONS: Uncertainty quantification can detect distribution shifts in radiotherapy auto-segmentation. This could support automated QA workflows by identifying in-distribution cases and enabling the application of specialised models.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Towards Trustworthy AI for Glioma Diagnosis: A Task-Aware Evaluation of Uncertainty QuantificationarXiv · 2026
- SpatialUQ: Post-Hoc Uncertainty Quantification from Spatial Consistency in Black-Box Vision ModelsarXiv · 2026
- Uncertainty as a Proxy for Semantic Correctness in Diffusion-Based Medical Image SynthesisarXiv · 2026
- OOD-SEG: Exploiting out-of-distribution detection techniques for learning image segmentation from sparse multi-class positive-only annotationsMedical Image Analysis · 2026
- Beyond Boundary Noise: Aggregated Aleatoric Uncertainty Fails to Capture Presence Ambiguity in 3D Lung Nodule SegmentationarXiv · 2026
- Improving confidence in MRI-based auto-segmentation via uncertainty assessmentActa Oncologica · 2026