Deep learning-based spot weight generation for online adaptive IMPT: implementation and technical feasibility on a clinical delivery system.
Bo Pang, Hong Quan, Tian Li, Heng Li, Minglei Kang, Lian Zhang, Yangguang Ma, Muyu Liu,
Cancer Center, Hubei Key Laboratory of Precision Radiation Oncology, Institute of Radiation Oncology, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, Hubei 430022, China; Key Laboratory of Biological Targeted Therapy (Huazhong University of Science and Technology), Ministry of Education, Wuhan, Hubei 430022, China. · Department of Medical Physics, School of Physics and Technology, Wuhan University, Wuhan 430072, China.
Key points
From the paper's conclusion- The proposed framework demonstrates the technical feasibility of generating IMPT spot weights with deep learning and implementing them on a clinical delivery system. Further improvement is needed for heterogeneous sites such as the lung.
Abstract
PURPOSE: To investigate the feasibility of a deep learning-based method for generating intensity-modulated proton therapy (IMPT) spot weights to enable a faster, more automated workflow. METHODS: A 3D Attention U-Net with a masked mean absolute error loss generated spot weights from four input channels (optimized plan dose, reference dose, CT image, spot mask), using planned spot weights as labels, and trained and tested on prostate, brain, and lung cases. Robustness was evaluated on target dose metrics only. The workflow was validated in the Eclipse® treatment planning system and delivered on a ProBeam 360° system. RESULTS: For weights in the range 0-5, mean absolute errors were below 0.8 for the prostate and brain models (15.9% and 48.1% of all weights, respectively) and below 1.7 for the lung model (31.2%). Mean relative errors were below 10% for the prostate model (5-65, 83.7%), 20% for the brain model (5-65, 51.7%), and 40% for the lung model (5-110, 68.6%); the remaining weights lay above these ranges and accounted for <0.5% of all weights. Gamma passing rates within the body contours (2%, 2 mm, 10% dose threshold) were 99.73 ± 0.32%, 98.95 ± 0.86%, and 92.75 ± 3.52% for the prostate, brain, and lung models, respectively. All delivered plans met the clinical QA acceptance criteria. CONCLUSION: The proposed framework demonstrates the technical feasibility of generating IMPT spot weights with deep learning and implementing them on a clinical delivery system. Further improvement is needed for heterogeneous sites such as the lung.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Applying a pretrained DL-based IMPT planning workflow in an external center: A feasibility study in oropharyngeal cancer.Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2026
- Optimization of Deep Learning Model for Denoising Dose Profiles in Proton TherapyarXiv · 2026
- A cascade deep learning network for simultaneous prediction of dose and linear energy transfer in proton therapy of prostate cancer.Physics in medicine and biology · 2026
- Noisy probing dose facilitated dose prediction for pencil beam scanning proton therapy: Physics enhances generalizabilityMedical Physics · 2026
- Quality assurance for deep learning-based proton therapy planning in routine clinical use for oropharyngeal cancer.Physics and imaging in radiation oncology · 2026
- Real world clinical outcomes of deep learning based versus manual proton therapy planning for oropharyngeal cancer: A matched cohort study.Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2026