Quantitative static contrast-enhanced phantom study to establish performance of advanced cone beam computed tomography for functional imaging applications.
Riley McMaster, Raanan Marants, Hedi Mohseni, Catherine Coolens
Department of Medical Biophysics, University of Toronto, Toronto, Ontario, Canada. · Princess Margaret Cancer Centre, University Health Network, Toronto, Ontario, Canada.
Key points
From the paper's conclusion- Advanced CBCT compared to standard CBCT improved image quality, approaching CT-level performance. Findings demonstrated robust volume and contrast-specific quantification under DCE conditions, supporting further evaluation of functional CBCT.
Abstract
BACKGROUND AND PURPOSE: Dynamic contrast-enhanced computed tomography (DCE-CT) is a quantitative imaging technique that measures vascular perfusion. Extending this to cone beam CT (CBCT) would enable functional measurements on radiotherapy machines. An advanced CBCT imaging solution, with enhanced image quality and rapid acquisitions, offers potential. However, the static quantitative and spatial performance of advanced CBCT relevant to DCE imaging has not been characterized. MATERIALS AND METHODS: To address this, the image quality of advanced CBCT was compared with standard CBCT and CT using contrast-enhanced phantoms with DCE-specific characteristics not previously applied to CBCT. A contrast enhancement CT (CECT) phantom (acrylic cylinder with capsule inserts) and a humanoid phantom (human torso with a liver containing tumour-mimicking spheres) were imaged across different acquisition settings. RESULTS: Using the CECT phantom, advanced CBCT demonstrated improved CT number accuracy compared to standard CBCT (<23.9 vs. <78.9 HU) and greater CT number constancy (<8.4 vs. <12.9 HU). Iterative reconstruction on advanced CBCT offered improvements over filtered back-projection. Using the humanoid phantom, findings were confirmed with CT number accuracy (<7.1 vs. <42.8 HU), CT number variation (<11.3 vs. <49.8 HU) and volume accuracy (<128.8 vs. <1067.3 mm3). Advanced CBCT achieved spatial resolutions of 2.3-5.0 lp/cm across different acquisitions and measurements, meeting the 5 lp/cm clinical benchmark under certain conditions. It also displayed stable CT number-density calibration. CONCLUSIONS: Advanced CBCT compared to standard CBCT improved image quality, approaching CT-level performance. Findings demonstrated robust volume and contrast-specific quantification under DCE conditions, supporting further evaluation of functional CBCT.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of quantitative accuracy in 4D DCE‐bCT: Insights from a perfusion phantom studyMedical Physics · 2026
- Evaluating Image Quality of Cone Beam Computed Tomography Using HyperSight® on the Varian TrueBeam.Annals of biomedical engineering · 2026
- Dual-energy cone-beam CT improves iodine contrast and Hounsfield unit value stability in the maxillofacial region: A phantom study.Biomedical physics & engineering express · 2026
- Towards adaptive radiotherapy: Comparing HyperSight and standard CBCT reconstructions on a C-Arm Linac.Journal of applied clinical medical physics · 2026
- Task-based image quality and dosimetric validation of HyperSight CBCT for adaptive radiotherapy.Journal of applied clinical medical physics · 2026
- A motion-compensated reconstruction algorithm improves reader-assessed cone-beam CT image quality during TACE.Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy · 2026