Department of Anesthesiology, Center for Pain Medicine, Erasmus MC University Medical Center, Rotterdam, The Netherlands. Electronic address: c.lansbergen@erasmusmc.nl. · Department of Neurology, The Migraine Clinic, Amsterdam, The Netherlands; Department of Neurology, Leiden University Medical Center, Leiden, The Netherlands.

Observational

Key points

From the paper's conclusion
  • Dedicated electrode leads for ONS are associated with a significantly lower subdistribution hazard ratio of lead-related revision surgery compared with conventional electrode leads, supporting their preferred use in ONS for MICCH.

Abstract

INTRODUCTION: Medically intractable chronic cluster headache (MICCH) is a highly disabling neurovascular brain disorder characterized by daily attacks of unilateral, excruciating headaches. Occipital nerve stimulation (ONS) can substantially reduce attack severity and frequency in patients with MICCH. However, its clinical application is limited due to high rates of lead migration and breakage requiring surgical revision. These complications are largely attributed to the initial off-label use of rigid electrode leads originally designed for spinal cord stimulation. To address these limitations, dedicated ONS electrode leads have been developed specifically for implantation in the occipital region. OBJECTIVES: To compare the risk of lead-related revision surgery between conventional and dedicated electrode leads for ONS in patients with MICCH. MATERIALS AND METHODS: We retrospectively included patients with MICCH who were treated with ONS at a single tertiary referral center between 2011 and 2021. Patients had received either a conventional off label spinal cord stimulation lead, a dedicated ONS lead, or both lead types sequentially after revision surgery. The primary end point was the risk of lead-related revision surgery (lead migration or breakage). Competing risk analysis accounted for nonlead-related revisions and explantations. RESULTS: We identified 115 patients treated with ONS, of whom 106 met all the inclusion criteria. In these patients, a total of 221 lead pairs were implanted (n = 95 conventional; n = 126 dedicated). The median follow-up was 2.8 years (interquartile range, 1.4-5.8) for conventional leads and 2.0 years (interquartile range, 0.8-3.1) for dedicated leads. In the competing risk analysis, dedicated leads were associated with a significantly lower risk of lead-related revision surgery compared with conventional leads (subdistribution hazard ratio, 0.144; 95% CI, 0.069-0.303; p < 0.001). Effectiveness was not formally evaluated, as a substantial proportion of the cohort overlapped with the ICON study population, in which ONS efficacy had previously been shown. CONCLUSIONS: Dedicated electrode leads for ONS are associated with a significantly lower subdistribution hazard ratio of lead-related revision surgery compared with conventional electrode leads, supporting their preferred use in ONS for MICCH.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers