Correlation between the IVIM-MRI based tissue perfusion parameters and blood flow index measured by diffuse correlation spectroscopy.
Neda Mogharari, Kamil Lipinski, Stanislaw Wojtkiewcz, Adam Liebert, Michal Kacprzak
Nalecz Institute of Biocybernetics and Biomedical Engineering, Polish Academy of Sciences, Poland. Electronic address: nmogharari@ibib.waw.pl. · Nalecz Institute of Biocybernetics and Biomedical Engineering, Polish Academy of Sciences, Poland; Institute of Radioelectronics and Multimedia Technology, Faculty of Electronics and Information Technology, Warsaw University of Technology, Poland.
Key points
From the paper's conclusion- Therefore, further studies with larger cohorts and different experimental protocols are required to better characterize the correlation between the IVIM- and DCS-derived parameters under varying perfusion conditions.
Abstract
Intravoxel incoherent motion (IVIM) magnetic resonance imaging (MRI) technique evaluates the microstructure and microvascular blood perfusion of biological tissues. IVIM-derived pseudo-diffusion coefficient (Dfast) quantifies the perfusion-related motion of blood within the tissue microvascular and is related to the mean squared displacement of water molecules in time. Diffuse correlation spectroscopy (DCS) is a noninvasive optical technique to monitor the microvascular blood flow changes. DCS-derived blood flow index (BFIDCS) is proportional to mean squared displacement of the red blood cells in time, thereby enabling comparison to Dfast. To evaluate the correlation between Dfast and BFIDCS, controlled muscle-occlusion tests were performed on five healthy volunteers, with five leg and four forearm datasets. In response to muscle occlusion, the BFIDCS decreased, whereas the mean Dfast increased, leading to no positive correlation between these parameters (rrm: repeated measures correlation, rrm=-0.219,95%CI:-0.724to0.438 in leg occlusion and rrm=-0.567,95%CI:-0.894to0.155, in forearm occlusion). The IVIM data were of adequate quality (median voxelwise signal to noise ratio at b = 0: 25-57), however the voxelwise coefficient of variation CVvoxels of the Dfast (median within-subjects CVvoxels in the leg: 210.3% and in forearm: 160.9%) were imprecise. The disagreement between the IVIM- and DCS-derived parameter responses is therefore specific to the applied protocol, and its cause remains unresolved. Therefore, further studies with larger cohorts and different experimental protocols are required to better characterize the correlation between the IVIM- and DCS-derived parameters under varying perfusion conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Towards Clinical Translation of Intravoxel Incoherent Motion MRI : Acquisition and Analysis Consensus RecommendationsJournal of Magnetic Resonance Imaging · 2026
- Assessing the fractional contributions of static, slow and fast dynamic scatterer components to the flow index derived by continuous wave diffuse correlation spectroscopybioRxiv · 2026
- Dynamic Macro-Microvascular Coupling During Cerebrovascular Reactivity Revealed by Simultaneous TCD and DCSmedRxiv · 2026
- Measurement of fluid diffusion coefficient based on bidirectional scanning in optical coherence tomography.Optics letters · 2026
- Evaluation of acquisition strategies in IVIM-DKI brain imaging: Effects on image quality, parameter stability, and comparison with DSC perfusion.Journal of medical imaging and radiation sciences · 2026
- Dynamic contrast-enhanced magnetic resonance imaging (MRI) and intravoxel incoherent motion imaging for characterizing angiogenesis in soft-tissue sarcomas: an exploratory study based on MRI-pathology co-alignment.Quantitative imaging in medicine and surgery · 2026