Assessing the repeatability and accuracy of quantitative double-echo steady-state (qDESS) as a wT2-mapping technique in test objects and the lower limb of healthy volunteers.
Gabriel Rossetto, David M Higgins, Kieren G Hollingsworth
Newcastle Magnetic Resonance Centre, Translational and Clinical Research Institute, Newcastle University, Newcastle Upon Tyne, United Kingdom. Electronic address: gabriel.rossetto@newcastle.ac.uk. · Clinical Science, Philips, Farnborough, United Kingdom.
Key points
From the paper's conclusion- 3D qDESS produces repeatable and accurate wT2 maps of the lower limb with substantially shorter acquisition times than the MESE sequence, making it a viable tool for clinical trials of MDs.
Abstract
PURPOSE: A multiple-echo spin-echo (MESE) acquisition is conventional for T2-mapping of skeletal muscle. Its time-intensive nature can make it impractically long for muscular dystrophies (MDs) such as facioscapulohumeral muscular dystrophy (FSHD). Here we assess the feasibility of using 3D quantitative double-echo steady-state (qDESS) to quantify muscle-water T2 (wT2) by evaluating the repeatability and accuracy of the estimated wT2 values. METHODS: First, the precision of qDESS as a wT2-mapping technique was determined by comparing against MESE and stimulated echo acquisition mode (STEAM) wT2 values from a phantom of eight MnCl2-doped water phantoms with concentrations ranging from 0.1-0.45 mM in 0.05 mM increments. Then, test-retest qDESS muscle-water wT2 maps of six lower limb muscles from eight healthy volunteers underwent a Bland-Altman analysis to assess repeatability. qDESS- and MESE-derived wT2 values were assessed against STEAM wT2 values at matched locations along the gastrocnemius medialis of a further five healthy volunteers to assess and compare accuracy. RESULTS: qDESS showed good repeatability and strong agreement with STEAM in test objects and the lower limb. CONCLUSION: 3D qDESS produces repeatable and accurate wT2 maps of the lower limb with substantially shorter acquisition times than the MESE sequence, making it a viable tool for clinical trials of MDs.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- T2 mapping of individual fascicles of the sciatic nerve at 7 T using multi-echo spin–echoMagnetic Resonance Materials in Physics Biology and Medicine · 2026
- Whole-body quantitative muscle MRI in myotonic dystrophy type 2: a sensitive tool for pattern recognition, disease stratification and objective assessment of clinical severity.Journal of neurology, neurosurgery, and psychiatry · 2026
- Quantitative MRI Comparison of Lower Limb Skeletal Muscle Water T 2 in Adults With and Without Probable SarcopeniaJCSM Communications · 2026
- Dual-Echo bSSFP for Rapid T2-Sensitive Quantitative Contrast Imaging at Ultra-Low Field.IEEE transactions on bio-medical engineering · 2026
- Textural Features from Quantitative MRI Using T2 Relaxation Times Correlate with Histological Cartilage Degeneration: An Ex-Vivo Study.Cartilage · 2026
- Combined qualitative and quantitative MRI reveals short-term progression in facioscapulohumeral muscular dystrophy.Journal of neuromuscular diseases · 2026