Journal of microbiological methods

Clinical performance, operational efficiency and turnaround time: A comparison of batch-based (cobas® respiratory flex) and on-demand (BIOFIRE® respiratory 2.1 panel) respiratory testing platforms.

Seok Hwee Koo, Boran Jiang, Zipeng Zhao, Tiffany Sze Min Low, Pei Qi Lim, Thean Yen Tan

Research Laboratory, Changi General Hospital, 2 Simei Street 3, Singapore 529889, Singapore. Electronic address: seok_hwee_koo@cgh.com.sg. · Department of Laboratory Medicine, Changi General Hospital, 2 Simei Street 3, Singapore 529889, Singapore. Electronic address: jiang_boran@cgh.com.sg.

Key points

From the paper's conclusion
  • Median turnaround time (TAT) was 1 h 10 min for RP2.1 versus 13 h 29 min for cobas®, with the longer TAT largely attributable to batch scheduling rather than analytical run time.

Abstract

The cobas® Respiratory flex assay was evaluated against the BIOFIRE® Respiratory 2.1 Panel (RP2.1) using 218 nasopharyngeal specimens. Cobas® achieved 100% PPA/NPA for influenza A/B, RSV, SARS-CoV-2, human metapneumovirus, and parainfluenza virus type 1, and failed to detect any of the four adenovirus-positive specimens identified by RP2.1 (PPA 0%, 95% CI: 0-49%); all estimates were based on small positive target counts and confidence intervals were wide. All four adenovirus-positive specimens were also positive by Allplex™. Median turnaround time (TAT) was 1 h 10 min for RP2.1 versus 13 h 29 min for cobas®, with the longer TAT largely attributable to batch scheduling rather than analytical run time. Cobas® required lower hands-on time at daily processing volumes exceeding four to ten samples.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers