Resolving Alinity m multi-Collect media incompatibility with guanidinium-based lysis methods used for Neisseria gonorrhoeae supplemental testing.
Todd M Pryce, Jan Williamson
PathWest Laboratory Medicine WA, Department of Clinical Microbiology, Fiona Stanley Hospital, Murdoch, Western Australia, Australia. Electronic address: todd.pryce@health.wa.gov.au. · Molecular Medicine, Royal Hobart Hospital, Hobart, Tasmania, Australia.
Key points
From the paper's conclusion- These findings demonstrate that the guanidine hydrochloride pre-extraction method effectively removes precipitate from Alinity media and enables reliable supplemental NG testing, with performance equivalent to cobas PCR media across both extraction-based and sample-to-answer platforms.
Abstract
False-positive Neisseria gonorrhoeae (NG) screening results require supplemental nucleic acid amplification testing for confirmation; however, Alinity m multi-Collect media (Alinity media) forms a precipitate when combined with guanidinium-based lysis buffers, interfering with nucleic acid extraction and causing Xpert CT/NG assay failures. This study evaluated a simple pre-extraction method to enable supplemental NG testing from Alinity media. A guanidine hydrochloride pre-extraction step was developed to remove precipitate prior to testing. Sixteen NG-positive urine specimens and a swab dilution series were inoculated into both Alinity media and cobas PCR media as a control. Samples were tested using Xpert CT/NG (NG2 and NG4 targets) and extracted on the MagNA Pure 96 platform for supplemental testing with ResistancePlus GC (RP-GC), targeting opa and porA. Qualitative results from Alinity media and cobas PCR media were in complete agreement, with paired quantification cycle (Cq) values showing high concordance across all targets and internal controls. RP-GC opa and porA targets demonstrated mean Cq differences of -0.45 and - 0.47, respectively, indicating a small but statistically significant earlier Cq for Alinity media (p = 0.004 and p = 0.018). Despite reaching statistical significance, these differences were less than 0.5 cycles and were not considered analytically meaningful. Xpert NG2 and NG4 targets showed negligible, non-significant Cq shifts. These findings demonstrate that the guanidine hydrochloride pre-extraction method effectively removes precipitate from Alinity media and enables reliable supplemental NG testing, with performance equivalent to cobas PCR media across both extraction-based and sample-to-answer platforms.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-104 Compatibility of Copan UBM Transport Medium with Molecular Detection of STI Targets across Different qPCR AssaysClinical Chemistry · 2026
- A large audit of Cobas CT/NG screening and Xpert CT/NG supplemental testing reveals low confirmatory rates in oropharyngeal samples with evidence of commensal cross-reactivity with an isolate of Neisseria lactamica.Epidemiology and infection · 2026
- Development and Laboratory Evaluation of a porA-Targeting TaqMan qPCR Assay for Neisseria gonorrhoeae: Clinical Specimen Validation and Molecular Estimation Using Cultured Reference Strains.International journal of molecular sciences · 2026
- Molecular typing and characterisation of antimicrobial resistance determinants of Neisseria gonorrhoeae among symptomatic individuals in the Eastern Cape, South AfricamedRxiv · 2026
- Comparison of Disk Diffusion and Agar Dilution for Neisseria gonorrhoeae Susceptibility Testing and Update.Clinical laboratory · 2026
- A-365 A Rapid Multiplex LAMP Assay for Detection of Chlamydia trachomatis, Neisseria gonorrhoeae, Trichomonas vaginalis, and Fluoroquinolone Resistance in Neisseria gonorrhoeae with Translatability Across Benchtop and Point-of-Care PlatformsClinical Chemistry · 2026