Analytical Validation of MethyLiverDx-Sigma, a Plasma cfDNA Methylation Assay for Hepatocellular Carcinoma Detection.
Si-Cho Kim, Hyunwoo Choi, Jiin Chung, Seung-Taek Lee, Young-Joon Kim
R&D Center, LepiDyne Inc., Seoul, Republic of Korea. · R&D Center, LepiDyne Inc., Seoul, Republic of Korea; Interdisciplinary Program of Integrated OMICS for Biomedical Science, Yonsei University, Seoul, Republic of Korea.
Key points
From the paper's conclusion- These findings characterize the analytical precision and multisite reproducibility of the PCR-HRM detection module and the analytical specificity of MLDx-s, complementing its previously reported clinical performance for plasma-based HCC detection.
Abstract
Hepatocellular carcinoma surveillance relies on ultrasonography and alpha-fetoprotein, both limited in sensitivity for early-stage disease; aberrant methylation of circulating cell-free DNA (cfDNA) offers a promising blood-based approach. Reported here is the analytical validation of MethyLiverDx-Sigma (MLDx-s), an assay that detects methylated RNF135 and DNMT3A in bisulfite-converted plasma cfDNA by methylation-specific PCR with high-resolution melting (HRM), with GAPDH as an endogenous control. Performance was evaluated following CLSI EP05-A3, EP07, EP12-A2, and EP37. Precision of quantification cycle (Cq) and melting peak temperature (Tm) was assessed by variance-component analysis within one site and across two sites and two operators. Because precision studies used bulk bisulfite-converted reference DNA, the reported coefficients of variation (CVs) primarily characterize the PCR-HRM detection module; end-to-end variability in clinical plasma specimens may be higher. Interference and cross-reactivity were evaluated in low-positive and contrived-negative samples against reproducibility-derived limits, with dose-response testing of reactive substances. Detection was 100% at both levels, and every run met control-validity criteria. Cq repeatability CVs ranged from 0.53% to 1.68%, and total reproducibility CVs ranged from 1.10% to 2.37%; all Tm CVs were ≤0.20%. All 18 interferents and 16 microbial cross-reactants evaluated in low-positive samples met the predefined acceptance criteria. In negative samples, hemoglobin, SARS-CoV-2, and measles virus produced concentration-dependent non-specific amplification, and dose-response testing established their respective no-effect concentrations. These findings characterize the analytical precision and multisite reproducibility of the PCR-HRM detection module and the analytical specificity of MLDx-s, complementing its previously reported clinical performance for plasma-based HCC detection.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The genomic and epigenomic abnormalities of plasma cfDNA as liquid biopsy biomarkers to detect hepatocellular carcinoma: a multicenter cohort studyJournal of Hematology & Oncology · 2025
- A multi‐centre, prospective trial of a methylation‐based liquid biopsy for early detection of liver cancer in high‐risk populationsClinical and Translational Medicine · 2026
- Diagnostic Performance of cfDNA-Based Liquid Biopsy for Hepatocellular Carcinoma: A Systematic Review and Meta-Analysis.Cancer science · 2026
- A multi-analyte cfDNA-based blood test for early detection of hepatocellular carcinomaJournal of Hepatology · 2026
- Cell-free DNA methylation markers in the diagnosis of hepatitis B virus-related- hepatocellular carcinoma: a review.Molecular biology reports · 2026
- Panel-level multilocus methylation quantification in native cell-free DNA by PCR-compatible sequential enzymatic processingmedRxiv · 2026