Journal of virological methods

Diagnostic efficacy of XQ-Station POCT, real-time PCR, and Droplet Digital PCR in detecting African swine fever virus.

Thi Ngoc Tram Ngo, Phuong Nguyen Le Dinh, Doo-Sung Cheon, Hyemin Park, Hoekyu Lee, Jinsun Kim, Nghia Mai Chi, Duy Bao Dung Nguyen,

Nong Lam University. Electronic address: tramngo.vsrt@gmail.com. · Nong Lam University. Electronic address: nldphuong1710@gmail.com.

Key points

From the paper's conclusion
  • This study aimed to compare the diagnostic performance of the XQ-Station point-of-care real-time PCR (POCT) system and p72 droplet digital PCR (ddPCR) for African swine fever virus (ASFV) detection, using the WOAH-recommended p72 real-time PCR (qPCR) assay as the reference.

Abstract

This study aimed to compare the diagnostic performance of the XQ-Station point-of-care real-time PCR (POCT) system and p72 droplet digital PCR (ddPCR) for African swine fever virus (ASFV) detection, using the WOAH-recommended p72 real-time PCR (qPCR) assay as the reference. A total of 132 clinical samples from 45 pigs, including oropharyngeal-nasal swabs, whole blood, and spleen tissues, were tested in parallel. Among 96 samples from ASFV-positive farms, qPCR, POCT, and ddPCR detected 83 (86.5%), 81 (84.4%), and 93 (96.9%) positive samples, respectively, while all 36 samples from an ASFV-negative farm tested negative. Compared with qPCR, POCT showed high agreement (κ = 0.968), relative sensitivity of 97.6%, specificity of 100%, and accuracy of 98.5%, with no significant difference in qualitative classification (p > 0.05). In contrast, ddPCR identified 10 additional positive samples, with substantial agreement (κ = 0.831; p < 0.01). Both assays effectively discriminated qPCR-positive and -negative samples (AUC = 0.988 and 0.984, respectively). Low viral DNA loads contributed to inter-platform discrepancies. Despite significant correlations, Bland-Altman analysis indicated insufficient quantitative agreement for interchangeable viral load quantification. POCT offers rapid field detection, whereas ddPCR provides complementary sensitivity for low viral DNA loads.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers