Journal of the Society for Cardiovascular Angiography & Interventions

Balloon-Assisted Countertraction Through the Central Fenestration for Successful Release of a Novel Fenestrated Atrial Septostomy Device: A Case Report.

Nithesh Kumar, Manokar Panchanatham, Rajkumar Ramalai Annamalai, Jebaraj Rathinasamy, Sundar Chidambaram

Department of Cardiology, Kauvery Hospital, Tamil Nadu, India.

Case report

Key points

From the paper's conclusion
  • We report an 85-year-old man who underwent septostomy using a novel custom fenestrated interatrial shunt device (ThoroughFare; Meril Life Sciences).

Abstract

Atrial septostomy is a palliative option for selected patients with severe pulmonary arterial hypertension and refractory right heart failure. We report an 85-year-old man who underwent septostomy using a novel custom fenestrated interatrial shunt device (ThoroughFare; Meril Life Sciences). Although device positioning was satisfactory, release from the delivery cable failed despite repeated attempts and snare-assisted stabilization. As a bailout strategy, a guidewire was advanced through the central fenestration and exchanged for an Amplatz stiff wire, followed by inflation of a 5 × 40 mm balloon within the fenestration to provide countertraction. This enabled successful device release without displacement. Balloon-assisted central fenestration stabilization may represent a useful bailout technique for challenging release scenarios in fenestrated structural heart interventions.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers