A single-run LC-MS/MS method for six parent statins and eight metabolites in human serum: Isomer baseline separation and clinical application.
Ziyun Li, Zixing Cai, Xinyue Wang, Hongna Mu, Hongxia Li, Siming Wang, Jiang Peng, Mingyang Hao,
Department of Basic Innovation Research, Beijing Hospital, National Center for Gerontology National Clinical Research Center for Gerontology; The Key Laboratory of Geriatrics of NHC; Beijing Key Laboratory of Aging Mechanism and Intervention Research on Aging-Related Diseases; Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing, 100730, PR China. · Department of Cardiology, Beijing Hospital, National Center of Gerontology; Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing, 100730, PR China.
Key points
From the paper's conclusion- This study aimed to develop a high-throughput LC-MS/MS method for 6 commonly prescribed statins and 8 metabolites in human serum.
Abstract
Statins are first-line lipid-lowering agents for cardiovascular disease, yet poor adherence and high inter-individual metabolic variability limit their clinical efficacy. This study aimed to develop a high-throughput LC-MS/MS method for 6 commonly prescribed statins and 8 metabolites in human serum. Ethyl acetate liquid-liquid extraction was performed on 100 μL of serum. A pentafluorophenyl column achieved baseline separation of two atorvastatin isomer pairs, with detection via multiple reaction monitoring and rapid polarity switching. Full bioanalytical validation was conducted, and 891 coronary angiography serum samples were analyzed for prescription consistency, dose-concentration relationships, and associated influencing factors. All 14 analytes were separated within 5 min and met standard criteria: linearity (r > 0.998), intra- and inter-batch imprecision of 1.0%-13.5%, and relative recoveries of 93.7%-116.7%. No significant carryover was observed, and all analytes were stable at -80 °C for 30 d. A consistency rate of 88.2% with prescription records confirmed large-scale applicability. Dose-dependent increases in serum concentration were observed for atorvastatin (P for trend < 0.01) and pitavastatin (P < 0.05). Active atorvastatin metabolites concentrations correlated negatively with blood lipids. Serum concentrations of simvastatin, rosuvastatin, and pravastatin were positively correlated with age. Correlations with metabolic and hemodynamic indices were compound-specific. The non-significant results for pitavastatin and fluvastatin likely reflected limited subgroup sizes. This single-injection method reliably quantifies 14 analytes. Its key strengths include isomer baseline separation, one-injection coverage of mainstream statins, and suitability for large cohorts, showing potential to support real-world statin exposure assessment, medication adherence verification, and routine therapeutic drug monitoring.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Implementation of a Pharmacogenetic Strategy for Statin Therapy in the Spanish National Health System: A Systematic Review and Regulatory Framework Analysis.Pharmaceuticals (Basel, Switzerland) · 2026
- Lipoprotein(a) variability in high CV risk patients with hypertriglyceridemia and controlled LDL-C on statin therapy – REDUCE-IT Lp(a) VariabilityEuropean Heart Journal · 2026
- Lipoprotein(a) levels in patients hospitalized for acute coronary syndrome according to prior statin therapyEuropean Heart Journal · 2026
- Variability in lipoprotein(a) measurements in ODYSSEY OUTCOMES: is one-and-done testing sufficient?European Heart Journal · 2026
- Real world evaluation of lipoprotein (a) testing and impact on LDL-C management in atherosclerotic cardiovascular disease patients: findings from the CVMOBIUS2 registryEuropean Heart Journal · 2026
- Predicting LDL-cholesterol reduction variability on statin-ezetimibe therapy: insights from "Jena auf Ziel".Herz · 2026