科研速览 · Science Skim继续刷下去 · Keep skimming →
◆ Food chemistry: X2026-08-01

Dual-mode screening system for rapid and accurate aflatoxin B1 detection in edible oils.

Guangyao Ying, Tingting Wang, Kunlun Li, Zheng Lu, Yuxin Wang, Gangjian Lin, Jun Li, Huili Xia, Liang Hong

原始摘要(英文原文)· Original abstract
This study established a dual-mode system integrating a portable flow cytometer with UHPLC-QqQ MS/MS for rapid AFB1 detection in edible oils. A competitive immunofluorescence sensor used streptavidin-magnetic microspheres with biotin-AFB1-BSA, AFB1 antibody, and APC-labeled secondary antibody. The method showed a linear range of 0.025-25 ng/g (R2 = 0.9966), with IC10 of 0.26 ng/g and IC50 of 0.76 ng/g. Recoveries were 90.51-102.73% with precisions ≤9.54%. UHPLC-QqQ MS/MS achieved LODs of 0.20-0.40 ng/g and LOQs of 0.50-1.25 ng/g for AFB1, AFB2, AFG1, and AFG2, with recoveries of 88.99-107.12% and precisions of 1.09-10.58%. Validation with 29 commercial oils and 10 contaminated peanut batches found no AFB1 in market samples and 1.07-9.36 ng/g in contaminated ones, consistent with UHPLC-QqQ MS/MS. The system enables on-site high-throughput screening with a flow cytometric reading speed of approximately 1 min per sample and a batch throughput of ∼32 samples/h (96-well plate format), while maintaining laboratory-grade accuracy. By enabling efficient triage of large sample volumes at the point of sampling, this strategy provides a practical and cost-effective framework for strengthening aflatoxin surveillance in the edible oil supply chain.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Dual-mode screening system for rapid and accurate aflatoxin B1 detection in edible oils. — 科研速览 Science Skim