Accuracy of intraoperative image-free and tracker-free mixed reality-guided pedicle screw placement across the entire spine: a cadaveric study.
Masayuki Ohashi, Masayuki Sato, Hideki Tashi, Yohei Shibuya, Keitaro Minato, Tsuyoshi Arabiki, Mio Kubota, Hiroyuki Sekimoto,
Division of Orthopaedic Surgery, Department of Regenerative and Transplant Medicine, Niigata University Graduate School of Medical and Dental Sciences, Niigata City, Japan. Electronic address: masayuki-ohashi@ksh.biglobe.ne.jp. · Division of Orthopaedic Surgery, Department of Regenerative and Transplant Medicine, Niigata University Graduate School of Medical and Dental Sciences, Niigata City, Japan.
Key points
From the paper's conclusion- These findings suggest that this simplified image-free and tracker-free MR navigation technique may offer a feasible approach for guiding spinal instrumentation.
Abstract
To evaluate the accuracy of intraoperative image-free and tracker-free mixed reality (MR)-guided pedicle screw (PS) placement across the entire spine using human cadavers. Eight human cadaveric specimens were used in this study. Three-dimensional holograms of the cadavers were automatically generated from preoperative CT data uploaded to the Holoeyes MD service (Holoeyes Inc., Tokyo, Japan). Using a Magic Leap One® headset (Magic Leap Inc., Plantation, FL, USA), the hologram with predetermined PS trajectory lines was superimposed onto the cadaveric spine, and PSs were inserted following the trajectory lines from C2 to L5. Accuracy was evaluated using the Gertzbein-Robbins classification: Grade A (screw within the pedicle), B (breach < 2 mm), C (2-4 mm), D (4-6 mm), and E (>6 mm). Grades A and B were considered clinically acceptable. Of the 261 PSs placed (93 cervical, 118 thoracic, and 50 lumbar), 166 (63.6%) were classified as Grade A, 88 (33.7%) as Grade B, 5 (1.9%) as Grade C, 2 (0.8%) as Grade D, and none as Grade E, resulting in an overall clinically acceptable placement rate of 97.3%. The clinically acceptable placement rate differed significantly among spinal regions (p = 0.030), with the lowest rate observed in the cervical spine (93.5%), compared with the thoracic (99.2%) and lumbar (100%) spine. This cadaveric study demonstrated that image-free and tracker-free MR-guided PS placement achieved a high rate of clinically acceptable screw placement. These findings suggest that this simplified image-free and tracker-free MR navigation technique may offer a feasible approach for guiding spinal instrumentation. However, the high accuracy achieved by a single experienced surgeon may represent a best-case scenario, and its generalizability to less experienced surgeons remains uncertain.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Augmented reality-enabled image-guided surgical navigation in minimally invasive spine surgery: application beyond screw placement in prone lateral transpsoas interbody fusion-a cadaveric surgical technique study.Journal of spine surgery (Hong Kong) · 2026
- New Mixed-Reality Navigation System for Cup Positioning in Trapeziometacarpal Arthroplasty: An In Vitro Validation Study.The journal of hand surgery Asian-Pacific volume · 2026
- Integrating mixed reality navigation in spine surgery: experience from a national spinal injuries unit and logistical-economic analysis.Irish journal of medical science · 2026
- Robotically equipped 3D hybrid operating room in spine and trauma surgery - updated experience in 359 patients and 2080 screws.Journal of robotic surgery · 2026
- A dual-modal fiducial marker with single-view x-ray pose recoverability for MR-guided surgical navigationScientific Reports · 2026
- Accuracy of Guide-Free, Mixed Reality-Assisted Patient Specific Implant Fixation for Maxillary Repositioning: A Cohort Study.Journal of dentistry · 2026