Towards Contact-Free Cardiac Magnetic Resonance: Feasibility of Remote Photoplethysmography for Cardiac Triggering and Gating.
Dr Med Patrick Doeblin, Karina Carrizosa, Jan Hendrik Wuelbern, Christian Stehning, Steffen Weiss, Stefanie Maria Werhahn, Ann-Sophie Eggers, Cyrill Meuwly,
Department of Cardiology, Angiology and Intensive Care Medicine, Deutsches Herzzentrum der Charité, Augustenburger Platz 1, 13353 Berlin, Germany; Charité - Universitätsmedizin Berlin, corporate member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Charitéplatz 1, 10117 Berlin, Germany; DZHK (German Centre for Cardiovascular Research), partner site Berlin, Potsdamer Str. 58, 10785 Berlin, Germany. Electronic address: patrick.doeblin@dhzc-charite.de. · Department of Cardiology, Angiology and Intensive Care Medicine, Deutsches Herzzentrum der Charité, Augustenburger Platz 1, 13353 Berlin, Germany; Charité - Universitätsmedizin Berlin, corporate member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Charitéplatz 1, 10117 Berlin, Germany.
Key points
From the paper's conclusion- Contact-free rPPG is a viable alternative to standard CMR triggering/gating using ECG or fPPG. Adjustment of our algorithm led to excellent trigger accuracy irrespective of skin type.
Abstract
BACKGROUND: Cardiac magnetic resonance (CMR) sequences necessitate synchronisation with the cardiac cycle. This is routinely obtained through electrocardiographic (ECG) detection of the R-wave with subsequent triggering or gating of the acquisition, depending on the type of sequence. An accepted alternative is the use of transmissive photoplethysmography using a fingerclip (fPPG) to detect the peripheral pulse wave. OBJECTIVES: To assess the feasibility of contact-free remote photoplethysmography (rPPG) for CMR triggering/gating. METHODS: In this prospective study (www.drks.de, DRKS00036686), CMR sequences were triggered/gated by ECG, fPPG and rPPG. The primary endpoint was trigger accuracy, measured by true positive rate (TPR), positive predictive value (PPV), beat detection error rate (BDER) and trigger-latency variability (TLV) in the rPPG and fPPG tracings compared to the ECG tracing. Secondary endpoints were technical success, image quality and quantitative CMR measurements of the acquired CMR sequences. Subsequently, the rPPG algorithm was adjusted to improve pulse wave detection in patients with skin type Fitzpatrick 5-6. RESULTS: 64 subjects were included. rPPG initially provided comparable measures of trigger accuracy to fPPG, with worse performance in subjects with skin type Fitzpatrick 5-6. Adjusting the rPPG algorithm significantly improved trigger accuracy irrespective of skin type: BDER (%) for fPPG 8.5 (0.9-23.0), for rPPG 7.2 (0.3-49.1) and for the adjusted rPPG 1.3 (0.2-11.2). Image quality with both fPPG and rPPG were comparable to ECG. The differences observed in quantitative CMR measurements between fPPG, rPPG and ECG were in the order of inter-observer-variability. CONCLUSIONS: Contact-free rPPG is a viable alternative to standard CMR triggering/gating using ECG or fPPG. Adjustment of our algorithm led to excellent trigger accuracy irrespective of skin type.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Camera-Seismocardiography Based Cardiac Triggering for Magnetic Resonance Imaging: A Prospective Study.IEEE journal of biomedical and health informatics · 2026
- Estimation of Contact PPG-Derived Vascular Indices from Reconstructed rPPG Signals.Sensors (Basel, Switzerland) · 2026
- Missed isochromatic cardiac pulsation in remote photoplethysmography: detection, impact, and removalBiomedical Optics Express · 2026
- Frequency-Preserving Readout Design and Error Characterization for CWT-Based rPPG Heart-Rate Estimation.Sensors (Basel, Switzerland) · 2026
- Feasibility of heart rate monitoring in hemodialysis patients using visible imaging: a pilot study.Frontiers in physiology · 2026
- SAMA: Training-Free Consensus Ridge Tracking for Camera-Based Heart Rate on Five Public DatasetsarXiv · 2026