Journal of magnetic resonance (San Diego, Calif. : 1997)

Diffusion imaging with a pulsed radiofrequency gradient spin-echo.

Millie A Barron, Charlotte R Sappo, Kevin D Harkins, Junzhong Xu, John C Gore, Jonathan B Martin

Institute of Imaging Science, Vanderbilt University, Nashville, TN, USA; Department of Physics and Astronomy, Vanderbilt University, Nashville, TN, USA. Electronic address: millie.a.barron@vanderbilt.edu. · Institute of Imaging Science, Vanderbilt University, Nashville, TN, USA; Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, USA; Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, TN, USA; Department of Chemical and Biomolecular Engineering, Vanderbilt University, Nashville, TN, USA; Department of Electrical and Computer Engineering, Vanderbilt University, Nashville, TN, USA.

Key points

From the paper's conclusion
  • The theory behind Bloch-Siegert RF-encoded diffusion imaging is described and b-values are derived based on estimates of the phase shifts imparted by off-resonant, inhomogeneous B1 pulses in place of conventional B0 diffusion-encoding gradients.

Abstract

We describe a novel MRI method for measuring and mapping apparent diffusion coefficients (ADCs) based on RF encoding. The theory behind Bloch-Siegert RF-encoded diffusion imaging is described and b-values are derived based on estimates of the phase shifts imparted by off-resonant, inhomogeneous B1 pulses in place of conventional B0 diffusion-encoding gradients. An RF-encoding coil with characteristic B1 variation along one axis was used to encode diffusion by the Bloch-Siegert shift in a modified PGSE sequence implemented on an ultra-low field 47.5-mT MRI system. ADC maps of phantoms containing aqueous solutions of polyvinylpyrrolidone were produced and ADC values were compared with those measured on a 3T system using conventional B0 diffusion encoding. Results show a similar PVP-concentration-dependent ADC trend between the novel technique and a typical diffusion-weighted imaging sequence. The potential advantages of Bloch-Siegert shift RF encoding for diffusion sensitization are discussed.

Read the paper