Pre-submission Pathological Quality Control for FoundationOne CDx Comprehensive Genomic Profiling.
Yume Natori, Yosuke Hirotsu, Masahiro Shibusawa, Kaito Nakamura, Shunsuke Watanabe, Saki Amemiya, Yuki Nagakubo, Ryosuke Tajiri,
Department of Pathology, Yamanashi Central Hospital, 1-1-1 Fujimi, Kofu, Yamanashi, Japan. · Genome Analysis Center, Yamanashi Central Hospital, 1-1-1 Fujimi, Kofu, Yamanashi, Japan. Electronic address: hirotsu-bdyu@ych.pref.yamanashi.jp.
Key points
From the paper's conclusion- These findings indicate most F1CDx assay failures are attributable to insufficient tumor content or specimen volume.
Abstract
Comprehensive genomic profiling (CGP) using FoundationOne CDx (F1CDx) requires formalin-fixed paraffin-embedded specimens with sufficient tumor area and DNA yield. Pre-submission pathological assessment is critical but has not been systematically evaluated in real-world practice. This study analyzed 371 consecutive F1CDx requests at a cancer genomic medicine designated hospital in Japan. Pre-submission quality checks assessed tumor surface area, estimated section numbers and residual specimen volume. Assay outcomes were classified as pre-submission cancellation, temporary suspension (insufficient DNA, suspected contamination, or technical issues), Qualified, or Pass. Of 371 requests, 52 (14.0%) were cancelled before submission, mainly due to insufficient specimen volume or clinical decision. Among 319 submitted cases, 22 (6.9%) were temporarily suspended. Among 297 cases with returned results, 256 (86.2%) achieved Pass quality and 41 (13.8%) were Qualified. Tumor content-related parameters including institutional tumor content, laboratory-assessed tumor content, and sequencing-estimated tumor content were all significantly lower in Qualified than in Pass case. In Qualified cases, evaluation of tumor mutational burden and microsatellite instability was successful in only 31.7% and 19.5%, respectively. In multivariable analysis, lower tumor content was the only independent predictor of QC failure. These findings indicate most F1CDx assay failures are attributable to insufficient tumor content or specimen volume. Strengthened pre-submission QC and coordinated workflows can meaningfully improve CGP success. Limitations include the retrospective single-institution design and lack of external validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-088 Comprehensive Genomic Profiling of Solid Tumors by Next Generation Sequence : Clinical Utility and Biomarker Discovery in a Pan-Cancer CohortClinical Chemistry · 2026
- Real-world clinical utility of comprehensive genomic profiling in advanced solid tumorsNature Medicine · 2026
- Pre-Analytical and Analytical Challenges in Whole-Exome Sequencing of Formalin-Fixed Paraffin-Embedded Breast and Prostate Cancer Tissue: A Real-World Multicenter StudyDiagnostics · 2026
- Association of sampling method and FFPE storage duration with comprehensive genomic profiling success in gastrointestinal tract cancersScientific Reports · 2026
- Platform Normalization and a Null Expectation as Preconditions for Reanalyzing Tumor Mutational Burden: A Reassessment of 3381 Breast Cancer Samples from Public Genomic CohortsInternational Journal of Environmental Research and Public Health · 2026
- Nationwide real‑world data on tumor‑agnostic genomic companion diagnostics and comprehensive genomic profiling testsMolecular and Clinical Oncology · 2026