Journal of the mechanical behavior of biomedical materials

Quantifying the impact of surface topography on the detectability of acoustic waves in hard tissue mimics using pump-probe microscopy.

Felix Wäger, Matthias Domke, Alexander Buchner, Heinz P Huber, Ronald Sroka

Laser-Forschungslabor, LIFE Center, University Hospital, LMU Munich, 82152 Planegg/Martinsried, Germany; Research Center for Microtechnology, FHV - Vorarlberg University of Applied Sciences, 6850 Dornbirn, Austria. Electronic address: felix.waeger@students.fhv.at. · Research Center for Microtechnology, FHV - Vorarlberg University of Applied Sciences, 6850 Dornbirn, Austria.

Key points

From the paper's conclusion
  • These findings demonstrate that optimal PPM sample preparation requires balancing pump-pulse energy transfer with speckle generation, showing that moderate surface roughness outperforms fine polishing in scattering media such as casein.

Abstract

Pump-probe microscopy (PPM) is a powerful technique for investigating laser-material interactions, such as the determination of material stiffness by detecting material-specific acoustic waves (AW). In scattering media, such as dentin or a matching artificial phantom, the detection of AWs relies on local wave-induced displacement of the surface speckle pattern, which results in signals often faint and difficult to distinguish from the background fluctuations. This study numerically models the underlying displacement effect and experimentally evaluates how surface topography influences AW detectability in homogeneous casein phantoms. Casein surfaces were systematically prepared using polishing grit sizes from 125 μm down to 1 μm (achieving Sq≈ 70 nm). Rough polishing scattered both pump-and probe-pulse, hindering AW generation and signal evaluation. Smoother surfaces enabled clear AW image acquisition, confirming theoretical simulation predictions. However, peak AW detectability occurred at a moderate surface roughness of Sq≈ 250 nm (5 μm grit). Contrary to the simulation, fine polished surfaces (1 μm grit) unexpectedly degraded image consistency and AW signals. These findings demonstrate that optimal PPM sample preparation requires balancing pump-pulse energy transfer with speckle generation, showing that moderate surface roughness outperforms fine polishing in scattering media such as casein. Furthermore, the calculated Young's modulus of casein 7.60 ± 0.12 GPa (mean ± SD) aligns with literature values for mammoth ivory and is slightly below natural human bone.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.