Blocker Eye, a novel video-enabled endobronchial blocker for one-lung ventilation during thoracic anesthesia: a porcine feasibility study.
Pak-Si Chan, Tsung-Ta Wu, Shun-Mao Yang, Chun-Yu Wu
Department of Anesthesiology, National Taiwan University Hospital Hsinchu Branch, Hsinchu, Taiwan. · Department of Surgery, National Taiwan University Hospital Hsinchu Branch, Hsinchu, Taiwan.
Key points
From the paper's conclusion- Blocker Eye enabled OLV with continuous endobronchial monitoring and maintained lung isolation during two resections. These preliminary findings support further device refinement and standardized preclinical safety evaluation before clinical testing.
Abstract
BACKGROUND: Conventional endobronchial blockers (EBBs) preserve a single-lumen airway but require fiberoptic bronchoscopy (FOB). Blocker Eye integrates imaging, illumination, steering, and continuous cuff-position monitoring. We evaluated its feasibility for porcine one-lung ventilation (OLV). METHODS: Three swine were studied. In Study 1, three board-certified anesthesiologists, all first-time in vivo users, performed 20 placements in one swine in consecutive device blocks: 10 with Blocker Eye and 10 with a conventional FOB-guided EBB. In Study 2, Blocker Eye provided lung isolation during wedge resection and segmentectomy in two swine. Placement time, procedural interruptions, physiology, and airway histopathology were assessed. RESULTS: In Study 1, OLV-establishment time was shorter with Blocker Eye (median [IQR], 30.9 [20.9-45.3] vs. 60.6 [50.9-71.0] s; exact Mann-Whitney U test, P = 0.009), although the comparison was exploratory and non-independent. In Study 2, both resections were completed, and the operating surgeon considered lung collapse sufficient. No ventilator interruption, rescue FOB, blocker exchange, repositioning, SpO2 <90%, or high airway pressure requiring intervention occurred; SpO2 remained 96-100% and end-tidal CO2 35-48 mmHg. Grade ≥2 epithelial loss occurred in 5/12 control, 11/12 single-application, and 9/12 repeatedly instrumented fields. Control-adjusted descriptive ISO 10993-6 framework scores were 4.0 for single application and 2.0 for repeated instrumentation. Injury was superficial with mild inflammation, but substantial epithelial loss in one uninstrumented bronchus precluded device-specific attribution. CONCLUSION: Blocker Eye enabled OLV with continuous endobronchial monitoring and maintained lung isolation during two resections. These preliminary findings support further device refinement and standardized preclinical safety evaluation before clinical testing.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The clinical trade-off of intraluminal versus extraluminal bronchial blockers in children: a retrospective analysis of respiratory mechanics and airway trauma during lung isolation preparationFrontiers in Pediatrics · 2026
- Fiberoptic Bronchoscopy-Guided Bronchial Blocker Placement in Lateral Versus Supine Positions in Video-Assisted Thoracic Surgery: A Comparative Study.Journal of visualized experiments : JoVE · 2026
- Closed Versus Open Vent Port Techniques and the Disconnection Technique for Optimizing Lung Collapse With Bronchial Blockers in Video-Assisted Thoracoscopic Surgery: A Randomized Controlled Trial.The clinical respiratory journal · 2026
- An Alternative Approach to Adult One-Lung Ventilation Using a Standard Single-Lumen Endotracheal Tube.Cureus · 2026
- Research progress in the application of double-lumen endobronchial tubes and bronchial blockers in thoracic surgery.Frontiers in medicine · 2026
- Nuances of perioperative lung isolation and airway management.Journal of thoracic disease · 2026