◆ Structural heart : the journal of the Heart Team2026-09-01
Durability Analysis of a Self-Expanding Transcatheter Aortic Valve Through 30 Years of Accelerated Wear Testing.
Guilherme Attizzani, Marco Frazzetto, Ole De Backer, Danny Dvir, Vanessa Garlepp, Tim Kelley, Kendra J Grubb, Gilbert H L Tang
原始摘要(英文原文)· Original abstract
•First reported 30-year in vitro accelerated wear testing data for a THV.•All Evolut PRO + valves met ISO hydrodynamic criteria through 20 years of testing.•At least one valve per size maintained performance through 30 years of testing.•Failures manifested in a consistent pattern, suggesting a shared mechanism.•Testing is ongoing until all valves fail.