Clinical amyloid subtyping from FFPE biopsies using a rapid preparation workflow and high-flow LC-Q Exactive MS.
Naga Veera Yerra, Srinivas V S Chakravartula, Coryn Gonis, Zead Elzoeiry, Aadhya Agarwal, Mohammad Zuber Nakadar, Thomas Hansen, Carlos Cordon-Cardo,
Clinical Chemistry Service, Department of Pathology, Molecular and Cell-Based Medicine, The Mount Sinai Hospital, New York, NY 10029, USA. Electronic address: nagaveeray@gmail.com. · Clinical Chemistry Service, Department of Pathology, Molecular and Cell-Based Medicine, The Mount Sinai Hospital, New York, NY 10029, USA; Department of Chemistry and Chemical Biology, Rutgers University, Piscataway, NJ 08854, USA.
Key points
From the paper's conclusion- We developed and evaluated a high-flow LC-MS/MS workflow for amyloid subtyping from formalin-fixed, paraffin-embedded (FFPE) tissue.
Abstract
Accurate amyloid protein typing is essential for diagnosis and treatment selection in patients with amyloidosis. Laser microdissection (LMD) followed by liquid chromatography-mass spectrometry provides characterization of amyloid deposits but may require prolonged sample preparation and analysis. We developed and evaluated a high-flow LC-MS/MS workflow for amyloid subtyping from formalin-fixed, paraffin-embedded (FFPE) tissue. Congo red-positive deposits were isolated by LMD from FFPE tissue and subjected to heat-assisted protein extraction and short-duration Trypsin/Lys-C digestion (3 h). Peptides were analyzed by high-flow LC-MS/MS using a Q Exactive Orbitrap with a 28-min gradient. Amyloid subtypes were identified on predefined protein sequence coverage, peptide-spectrum match, unique peptide, and amyloid-associated protein criteria. Analytical performance was evaluated using protein standards, LMD tissue dilution studies, precision, interference, stability, and blinded clinical specimens compared with a reference laboratory method. Sample preparation was completed within 6 h, enabling same-day analysis. A minimum LMD tissue area of >270,000 μm2 was established for reliable clinical interpretation. In blind clinical specimens, amyloid subtype assignment was concordant with Mayo Clinic reference laboratory results in 20 of 20 cases. Intra- and inter-run studies demonstrated stable chromatographic performance and reproducible amyloid protein identification. Increasing concentrations of albumin and hemoglobin reduced amyloid protein recovery, defining an important preanalytical limitation of the assay. This rapid high-flow LC-MS/MS workflow enables same-day amyloid protein subtyping from LMD FFPE tissue. The combination of standardized tissue input, short-duration digestion, short-gradient LC-MS/MS, and predefined protein identification criteria supports implementation of this approach in clinical laboratories.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Mass Spectrometry-Based Amyloid Typing in FFPE Biopsies: A Five-Year Prospective Experience from a Tertiary Academic Center in SpainmedRxiv · 2026
- Suspension trapping and data-independent acquisition enable high-flow liquid chromatography-tandem mass spectrometry-based amyloidosis typing in clinical laboratories.Clinical proteomics · 2026
- Mass Spectrometry-Based Diagnosis and Typing of Amyloidosis: Analytical Workflows, Interpretation and Clinical ImplementationTargets · 2026
- Reproducible Amyloid Typing by Data-Independent Acquisition Mass Spectrometry.Clinical chemistry · 2026
- Molecular Identification of Transthyretin Amyloid in Fixed Skin Biopsies: A Simplified Sample Preparation Method.Journal of visualized experiments : JoVE · 2026
- Detection, Monitoring, and Discrimination of Amyloidogenic vs Innocent Light Chains in Systemic Amyloidoses.Blood · 2026