Improving metabolite identification through enhanced positional isomer separation using micro-pillar array columns.
Arnaud Lubin, Peter Verboven, Filip Cuyckens
Translational PKPD and Investigative Toxicology, Johnson & Johnson, Beerse, Belgium. · Translational PKPD and Investigative Toxicology, Johnson & Johnson, Beerse, Belgium. Electronic address: FCUYCKEN@its.jnj.com.
Key points
From the paper's conclusion- These results demonstrate that μPAC/nano-ESI-MS represents a valuable workflow enhancement for metabolite identification studies and can be implemented as a second-pass analysis strategy when conventional UHPLC/ESI-MS methods provide insufficient separation power.
Abstract
Metabolite identification (MetID) workflows frequently rely on generic UHPLC/ESI-MS methods that provide a compromise between chromatographic separation and analytical throughput. However, these methods may fail to adequately resolve complex mixtures of metabolite isomers, resulting in ambiguous structural assignments and additional method development efforts. Here, we evaluate micro-pillar array columns (μPAC) as a practical strategy to improve isomer separation without changing chromatographic selectivity or developing dedicated LC methods. Using hepatocyte incubations with buspirone and verapamil as model compounds, μPAC/nano-ESI-MS was compared with conventional UHPLC/ESI-MS methods under both standard and extended-gradient conditions. For buspirone, μPAC/nano-ESI-MS substantially improved the separation of oxidative metabolite isomers and enabled detection of 14 out of 15 theoretically possible metabolites with two oxidations. For verapamil glucuronides, μPAC/nano-ESI-MS provided superior chromatographic resolution compared with both conventional UHPLC and a 90-min UHPLC gradient, demonstrating that the observed improvements extend beyond the effect of gradient length alone. In addition, μPAC/nano-ESI-MS produced more than an order-of-magnitude increase in signal response despite a 100-fold lower column load. The combined gains in chromatographic resolution and signal intensity improved confidence in metabolite annotation and facilitated the characterization of complex isomeric metabolite mixtures. These results demonstrate that μPAC/nano-ESI-MS represents a valuable workflow enhancement for metabolite identification studies and can be implemented as a second-pass analysis strategy when conventional UHPLC/ESI-MS methods provide insufficient separation power.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Systematic evaluation of SPE cartridges in a 12-second SPE-IM-MS workflow for untargeted intracellular metabolomicsScientific Reports · 2026
- Simultaneous enantiomeric separation and analysis of dimpropyridaz and its two major metabolites in plant-origin foods by UHPLC-MS/MS.Food chemistry · 2026
- DEPI-MS: Density-Equalized Pseudo-Image Learning for Drift-Tolerant LC-MS Metabolomics Classification.Analytical chemistry · 2026
- Enhancing metabolite coverage using dedicated mobile phases for individual polarity modes in HILIC-MSAnalytical and Bioanalytical Chemistry · 2025
- Sub-3 μm porous graphitic carbon for highly polar metabolomics: orthogonal selectivity and comparison with HILIC and reversed-phase chromatography.Analytical and bioanalytical chemistry · 2026
- Identifying Key Metabolites in South African Medicinal Plants Using Dual Electrospray Ionization MetabolomicsPlants · 2026