Safety and Efficacy of Transvenous Lead Extraction Using a Right-Sided Approach.
Joe Demian, Thomas Callahan, David O Martin, Arshneel Kochar, Pasquale Santangeli, Tyler Taigen, Medhat Farwati, Chadi Tabaja,
Department of Cardiovascular Medicine, Cardiac Electrophysiology Section, Cleveland Clinic, Cleveland, Ohio, USA. · Department of Cardiovascular Medicine, Cardiac Electrophysiology Section, Cleveland Clinic, Cleveland, Ohio, USA. Electronic address: younisa@ccf.org.
Key points
From the paper's conclusion- Right-sided TLE is associated with a lower rate of complete extraction and increased procedural risk, particularly sternotomy and major vascular injury, compared with left-sided TLE. These findings should be carefully weighed when implanting or extracting right-sided systems.
Abstract
BACKGROUND: Despite extensive experience with transvenous lead extraction (TLE) overall, outcomes for systems implanted on the right chest remain underexplored, largely because of their low incidence. OBJECTIVES: The objective was to evaluate whether system laterality influences TLE outcomes. METHODS: Using our institutional, prospectively maintained ExTRACT registry, we analyzed all consecutive patients undergoing TLE procedures at the Cleveland Clinic between 1996 and 2022. Patients were categorized by system laterality. Baseline characteristics, procedural characteristics, and complications were assessed. Multivariable logistic regression models evaluated the association between right-sided TLE and adverse events. RESULTS: Among 4,989 patients, 494 (9.9%) had right-sided and 4,495 (90.1%) had left-sided TLE. Compared with left-sided cases, right-sided cases had longer total lead dwell times (13.83 vs 9.26 years, P < 0.001) and were more likely to be extracted because of infection (47.57% vs 39.71%, P = 0.001). Fluoroscopy time was longer (15.95 vs 12.1 minutes; P < 0.001), and femoral workstation use more frequent (7.89% vs 4.05%, P < 0.001), yet complete extraction was lower (94.53% vs 96.44%, P = 0.034). Right-sided TLE had higher rates of sternotomy (2.83% vs 1.25%, P = 0.004) and superior vena cava or right atrial tear/perforation (2.49% vs 1.10%, P = 0.009). Thirty-day mortality was low and comparable (0.61% vs 0.47%, P = 0.727). Multivariable regression showed that right-sided TLE was associated with higher odds of sternotomy (OR = 1.92, 95% CI: 1.01-3.46, P = 0.037) and superior vena cava or right atrial tear/perforation (OR = 2.02, 95% CI: 1.01-3.78, P = 0.037). CONCLUSIONS: Right-sided TLE is associated with a lower rate of complete extraction and increased procedural risk, particularly sternotomy and major vascular injury, compared with left-sided TLE. These findings should be carefully weighed when implanting or extracting right-sided systems.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Safety and Efficacy of Right- Versus Left-Sided Transvenous Lead Extraction.Journal of cardiovascular electrophysiology · 2026
- Powered Rotational Sheaths in Complex Transvenous Lead Extraction: Comparative Outcomes With Telescoping Sheaths and the WARD-10 Score.Journal of cardiovascular electrophysiology · 2026
- Transvenous Lead Extraction Without Dedicated Surgical Back up in Low-Risk Patients.Heart rhythm · 2026
- Long-Term Mortality Following Transvenous Lead Extraction: Data from EXTRACT Registry.Journal of clinical medicine · 2026
- Safety and Outcomes of Transvenous Lead Extraction of Advisory ePTFE-Coated Implantable Cardioverter-Defibrillator Leads: A Multicenter Experience With RELIANCE G.JACC. Clinical electrophysiology · 2026
- Transvenous Lead Extraction Using a Stepwise Bailout Strategy: A Single-Center Experience with Wrap-Around Rescue in Selected Bailout Cases.Journal of clinical medicine · 2026