JACC. Case reports

Percutaneous Multivalvular Endocarditis Debulking With a Whole-Body Embolic Protection Device.

Ahmad Younes, Carson Oostra, Robert Roether, Donald Crescenzo, Mohamad Dabaja, Claudiu Georgescu, Stephanie Younes

Jobst Vascular Institute, Promedica, Toledo, Ohio, USA. · Promedica Physicians Cardiology, Promedica, Toledo, Ohio, USA.

Key points

From the paper's conclusion
  • This represents, to our knowledge, the first reported use of whole-body embolic protection during aortic and combined tricuspid/mitral percutaneous endocarditis debulking.

Abstract

BACKGROUND: Percutaneous debulking of left-sided endocarditis and thrombus has previously relied on cerebral-only embolic protection devices. The Emboliner is an investigational nitinol-mesh filter that provides whole-body embolic protection and has been studied in transcatheter aortic valve replacement. SUMMARY: We describe its first reported use as an adjunct to AngioVac-based debulking in 2 patients: 1 with aortic valve endocarditis and 1 with combined tricuspid and mitral endocarditis. DISCUSSION: The Emboliner was successfully deployed in both cases, with over 90% of vegetation removed and no periprocedural or in-hospital embolic events. Deployment technique differed by vegetation location: in the aortic valve case, the AngioVac cannula was advanced through the filter's access port; in the tricuspid/mitral case, no further interaction with the filter was required once deployed. NOVELTY: This represents, to our knowledge, the first reported use of whole-body embolic protection during aortic and combined tricuspid/mitral percutaneous endocarditis debulking. TAKE-HOME MESSAGE: Whole-body embolic protection may offer an important adjunct to percutaneous debulking in patients at high surgical risk.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers