Mechanical Hemolysis After Fontan Fenestration Closure: An Unrecognized Complication.
Mohammad Abed, Nicholas Fitzgerald
Division of Cardiology, Labatt Family Heart Center, The Hospital for Sick Children, Temerty School of Medicine, Toronto, Ontario, Canada. Electronic address: Mohammad.abed@sickkids.ca. · Division of Cardiology, Labatt Family Heart Center, The Hospital for Sick Children, Temerty School of Medicine, Toronto, Ontario, Canada.
Key points
From the paper's conclusion- Hematuria after Fontan fenestration closure should raise suspicion for mechanical hemolysis, even without residual shunt.
Abstract
BACKGROUND: Mechanical hemolysis following transcatheter device closure is uncommon and is typically associated with a residual high-velocity shunt. Hemolysis after Fontan fenestration closure has not been previously reported. CASE SUMMARY: A 7-year-old boy with single-ventricle physiology underwent elective transcatheter Fontan fenestration closure, complicated by hematuria the following day. Laboratory evaluation confirmed nonimmune intravascular hemolysis. Echocardiography demonstrated a well-positioned device without residual shunt but with an atrioventricular valve regurgitant jet directed toward the device. Conservative management was pursued, with biochemical surveillance, folic acid, and iron supplementation. Device removal was not required. DISCUSSION: Device-related hemolysis is classically attributed to residual shunting. Described herein is hemolysis caused by regurgitant jet collision with a nonendothelialized device. Conservative management is appropriate in stable patients, as given expected resolution with device endothelialization. TAKE-HOME MESSAGES: Hematuria after Fontan fenestration closure should raise suspicion for mechanical hemolysis, even without residual shunt. Early recognition is essential in Fontan patients and warrants earlier supportive intervention given limited cardiopulmonary reserve.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Post Atrial Septal Defect Device Closure Acute Severe Haemolysis: one hole closed, one valve leaked-A Rare Case Report.Cardiology in the young · 2026
- Mechanical haemolysis as a rare complication of modified Blalock–Thomas-Taussig shunt physiologyCardiology in the Young · 2026
- Successful transcatheter treatment of chronic mechanical haemolysis after cardiac surgery: a case report.Cardiology in the young · 2026
- Chorda Tendinea Obstruction of Impella Causing Massive Hemolysis Detected by Methemoglobinemia: A Case Report.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
- Clinical Implications of Hemolysis after Pulsed Field Ablation - What Can We Learn from Benchtop Models and Preclinical Experiments.Heart rhythm · 2026
- Hemolysis and Kidney Injury in Patients Receiving Mechanical Circulatory Support: A Mechanistic Review.Kidney360 · 2026