Recalibration of Invasive Hemodynamic Monitoring Devices on Serial Catheterization in Adults With Fontan Circulation.

Andrew M Freddo, Arash Salavitabar, Benjamin Blais, Aimee K Armstrong, Curt J Daniels, William H Marshall V

The Heart Center at Nationwide Children's Hospital, Columbus, Ohio, USA; Division of Cardiovascular Medicine, The Ohio State University Wexner Medical Center, Columbus, Ohio, USA. Electronic address: andrew.freddo@osumc.edu. · The Heart Center at Nationwide Children's Hospital, Columbus, Ohio, USA.

Key points

From the paper's conclusion
  • Because we find discrepancy in device measurements in our cohort, we recommend recalibration with each hemodynamic catheterization in this population.

Abstract

BACKGROUND: There are limited data on long-term accuracy of invasive hemodynamic monitoring (IHM) in patients with nonpulsatile Fontan circulation (FC). FIRST-IN-HUMAN/EARLY REPORTS SUMMARY: In a retrospective review of 20 patients with FC and CardioMEMS IHM implanted (2015-2025), simultaneous IHM and invasive mean pulmonary artery pressure measurements were available during 6 repeat catheterizations in 6 patients. IHM readings were nearly all lower than invasive measurements, with the absolute difference ranging from 1 to 10 mm Hg. DISCUSSION/NOVELTY: There are limited reports of recalibration data with IHM, and to our knowledge this is the first report of simultaneous IHM and invasive measurement in adults with FC. IHM accuracy is important with FC as small pressure differences can greatly affect physiology. TAKE-HOME MESSAGES: IHM can provide outpatient pulmonary artery pressure measurements in patients with FC. Because we find discrepancy in device measurements in our cohort, we recommend recalibration with each hemodynamic catheterization in this population.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers