Recalibration of Invasive Hemodynamic Monitoring Devices on Serial Catheterization in Adults With Fontan Circulation.
Andrew M Freddo, Arash Salavitabar, Benjamin Blais, Aimee K Armstrong, Curt J Daniels, William H Marshall V
The Heart Center at Nationwide Children's Hospital, Columbus, Ohio, USA; Division of Cardiovascular Medicine, The Ohio State University Wexner Medical Center, Columbus, Ohio, USA. Electronic address: andrew.freddo@osumc.edu. · The Heart Center at Nationwide Children's Hospital, Columbus, Ohio, USA.
Key points
From the paper's conclusion- Because we find discrepancy in device measurements in our cohort, we recommend recalibration with each hemodynamic catheterization in this population.
Abstract
BACKGROUND: There are limited data on long-term accuracy of invasive hemodynamic monitoring (IHM) in patients with nonpulsatile Fontan circulation (FC). FIRST-IN-HUMAN/EARLY REPORTS SUMMARY: In a retrospective review of 20 patients with FC and CardioMEMS IHM implanted (2015-2025), simultaneous IHM and invasive mean pulmonary artery pressure measurements were available during 6 repeat catheterizations in 6 patients. IHM readings were nearly all lower than invasive measurements, with the absolute difference ranging from 1 to 10 mm Hg. DISCUSSION/NOVELTY: There are limited reports of recalibration data with IHM, and to our knowledge this is the first report of simultaneous IHM and invasive measurement in adults with FC. IHM accuracy is important with FC as small pressure differences can greatly affect physiology. TAKE-HOME MESSAGES: IHM can provide outpatient pulmonary artery pressure measurements in patients with FC. Because we find discrepancy in device measurements in our cohort, we recommend recalibration with each hemodynamic catheterization in this population.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Invasive pulmonary artery pressure monitoring in heart failure: evidence, clinical application, and future directions.Heart failure reviews · 2026
- Comparing noninvasive blood pressure monitoring technologies: the challenge of between-arm differences.British journal of anaesthesia · 2026
- Role of digital health in the personalized management of heart failure due to Fontan failure: a case report.European heart journal. Case reports · 2026
- Six-year experience of CardioMEMS-guided nursing interventions in a multidisciplinary heart failure unitEuropean Heart Journal · 2026
- The map and the territory: a critical appraisal of pulse contour cardiac output monitoring.British journal of anaesthesia · 2026
- Accuracy of Indirect Fick Method for Systemic Flow Calculation In Adults With A Fontan Circulation.American heart journal · 2026