Leadless Transatrial or Transconduit Pacing in Fontan Circulation.
Rachel K Duncan, Mohammed Faluk, Angela J Weingarten, Benjamin P Frischhertz, Jamie Kowal, Majd El-Harasis, Travis Richardson, Dana Janssen,
Division of Cardiovascular Medicine, Vanderbilt University Medical Center, Nashville, Tennessee, USA. · Division of Cardiovascular Medicine, Vanderbilt University Medical Center, Nashville, Tennessee, USA; Division of Pediatric Cardiology, Monroe Carell Jr Children's Hospital at Vanderbilt, Nashville, Tennessee, USA.
Key points
From the paper's conclusion- This case series demonstrates leadless pacing in the pulmonary venous, atrial, and systemic ventricular chambers in Fontan circulation, including atrial AVEIR AR implantation and extended follow-up of ventricular Micra AV pacing. These approaches may expand pacing options in selected patients with Fontan when conventional strategies are limited, although long-term outcomes remain unknown.
Abstract
BACKGROUND: Epicardial or transvenous pacing strategies may be suboptimal in certain patients with Fontan requiring permanent pacing. Leadless pacing offers an alternative, but experience in complex Fontan anatomy remains limited. CASE SUMMARY: We report 3 patients with Fontan who underwent transatrial or transconduit leadless pacemaker (LP) implantation. One patient with a lateral tunnel Fontan and sinus node dysfunction underwent successful placement of an AVEIR AR (Abbott) LP in the pulmonary venous atrium. Two patients with Fontan circulation and complete heart block with prior epicardial pacing systems underwent implantation of a Micra AV (Medtronic) LP in the systemic ventricle, with longer follow-up than in prior reports. CONCLUSIONS: This case series demonstrates leadless pacing in the pulmonary venous, atrial, and systemic ventricular chambers in Fontan circulation, including atrial AVEIR AR implantation and extended follow-up of ventricular Micra AV pacing. These approaches may expand pacing options in selected patients with Fontan when conventional strategies are limited, although long-term outcomes remain unknown.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Transpulmonary Atrial Pacemaker Implantation in Fontan Circulation With Bilateral Glenn.JACC. Case reports · 2026
- Transvenous Atrial Pacing Using Subclavian 3-Dimensional Electroanatomical Mapping in Lateral Tunnel Fontan With Sinus Arrest.JACC. Case reports · 2026
- Novel Transvenous Approach for Implantable Cardioverter-Defibrillator Implantation in a Patient With Fontan Circulation: Trans-Left Pulmonary Artery-Left Atrium.JACC. Case reports · 2026
- Atrial leadless pacing for heart failure in sick sinus syndrome.Journal of cardiology cases · 2026
- Junctional Rhythm as Reversible Driver of Fontan Failure: Recovery With Multimodal Therapy Including Atrial Pacing.JACC. Case reports · 2026
- Transjugular implantation of a leadless pacemaker in a 97-year-old patient with severe kyphosis and tortuous femoral veins: a case report.Frontiers in cardiovascular medicine · 2026