Validating ECG-Based Artificial Intelligence Models for Cardiomyopathy Detection in Obstetric Patients: A Multisite International Study.
Demilade Adedinsewo, Patrick W Johnson, Carolina Morales-Lara, Maria Vitoria Coutinho, Sabrina D Phillips, Marc A Eidelhoch, Michal Cohen-Shelly, Elad Maor,
Department of Cardiovascular Medicine, Mayo Clinic, Jacksonville, Florida, USA. Electronic address: adedinsewo.demilade@mayo.edu. · Department of Quantitative Health Sciences, Mayo Clinic, Jacksonville, Florida, USA.
Key points
From the paper's conclusion- AI-ECG models demonstrated strong external validity in a global obstetric cohort, with significantly improved performance in the refined model trained on expanded data. These findings support the potential role of AI-ECG for a scalable LVSD screening tool in obstetric populations, while underscoring the importance of population-specific validation prior to clinical implementation.
Abstract
BACKGROUND: Cardiomyopathy significantly impacts maternal health worldwide. Artificial intelligence-enabled electrocardiograms (AI-ECG) have shown effectiveness in detecting left ventricular systolic dysfunction (LVSD) in adults and limited obstetric cohorts; however, broad validation in diverse obstetric populations remains unexplored. OBJECTIVES: Our primary objective was to validate the performance of 2 versions of an AI-ECG model for detecting cardiomyopathy in an international obstetric cohort. METHODS: We conducted a retrospective, multisite study of pregnant or postpartum women (≤12 months) across 11 institutions in 4 countries (Israel, Nigeria, Pakistan, and the United States). Two AI-ECG models (Food and Drug Administration-cleared v2.5.0 and refined v3.0.0) were evaluated against echocardiography. LVSD was defined as left ventricular ejection fraction ≤40% in keeping with the model's training target. Model performance was assessed by area under the receiver-operating characteristic curve (AUROC). RESULTS: We included 1,820 women with paired ECG-echocardiogram data, of whom 7.3% had LVSD. Model 1 (v2.5.0) demonstrated strong discriminative performance with an AUROC of 0.865, sensitivity of 46.1%, and negative predictive value of 95.8%. Model 2 (v3.0.0) showed significantly improved performance, with an AUROC of 0.925, sensitivity of 64.1%, and negative predictive value of 97.2% (P < 0.001) for LVSD detection. CONCLUSIONS: AI-ECG models demonstrated strong external validity in a global obstetric cohort, with significantly improved performance in the refined model trained on expanded data. These findings support the potential role of AI-ECG for a scalable LVSD screening tool in obstetric populations, while underscoring the importance of population-specific validation prior to clinical implementation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Artificial Intelligence-Enabled Electrocardiograms and Pregnancy-Related Cardiomyopathy: A Systematic Review and Diagnostic Meta-Analysis.JACC. Advances · 2026
- Artificial intelligence enabled electrocardiography and digital auscultation for detection of peripartum cardiomyopathy and pregnancy-associated left ventricular systolic dysfunction: A systematic review and meta-analysis of diagnostic accuracy studies.European journal of obstetrics, gynecology, and reproductive biology · 2026
- AI-enabled electrocardiography as a screening tool for peripartum cardiomyopathy: a meta-analysis of diagnostic accuracyEuropean Heart Journal · 2026
- Artificial Intelligence–Enhanced Electrocardiogram Models for Detection of Left Ventricular DysfunctionJACC Advances · 2026
- Utility of ECG-Vision for Surveillance of Left Ventricular Dysfunction in Patients With Hypertrophic Cardiomyopathy Initiated on Mavacamten.Circulation. Heart failure · 2026
- Prediction of Left Ventricular Systolic Dysfunction Using an Artificial Intelligence-Based Electrocardiogram Analysis Model in Patients Presenting to the Emergency Department.Diagnostics (Basel, Switzerland) · 2026