Rapid performance validation testing of an intraoperatively customizable antibiotic-eluting resorbable bone graft implant targeting prosthetic joint infection.
Eva Serio, Madeleine Gaztambide, David W Grainger, Ashok C Khandkar
Elute, Inc., 630 Komas Drive, # 200, Salt Lake City, Utah 84108 USA. · Elute, Inc., 630 Komas Drive, # 200, Salt Lake City, Utah 84108 USA; Departments of Biomedical Engineering and of Molecular Pharmaceutics, The University of Utah, Salt Lake City, Utah 84112 USA.
Key points
From the paper's conclusion- These findings produce a practical quality control assay for two FDA-cleared implant platforms, one currently in clinical trials for targeted treatment of PJIs and other orthopedic infections where infection control and bone healing must be simultaneously addressed locally.
Abstract
Periprosthetic joint infections (PJIs) present a major clinical challenge, for which prolonged, localized antibiotic delivery to eradicate infection are routinely pursued to facilitate bone regeneration while minimizing systemic toxicity. Current treatments, such as antibiotic-loaded polymethylmethacrylate bone cements, are non-resorbable and provide inconsistent antibiotic release. This study evaluates release of antibiotics from a novel, FDA-cleared, biphasic, synthetic, resorbable bone graft device composed of calcium carbonate and hydroxyapatite bound with polycaprolactone and polylactic-co-glycolic acid. This bone graft device can be intraoperatively mixed with patient-specific antibiotics (e.g., tobramycin and/or vancomycin) to form drug-loaded granules for implantation into bone voids. Drugs release completely from this formulation in 6-8 weeks. To rapidly screen and assess drug release consistency from these commercial formulations, a United States Pharmacopeia (USP) compendial dissolution method was employed, showing ∼ 100% tobramycin release within 145 min from this drug-loaded device. We then compared this rapid release to previously published actual 8-week release data and predictions from Weibull and Siepmann models fit to the published 8-week data. Results validate the utility of the rapid USP compendial dissolution method as a valuable, accelerated surrogate predictor for the previously published device-controlled antimicrobial release in vivo of over 6 weeks. These findings produce a practical quality control assay for two FDA-cleared implant platforms, one currently in clinical trials for targeted treatment of PJIs and other orthopedic infections where infection control and bone healing must be simultaneously addressed locally. This approach facilitates performance verification of an intraoperatively customizable, antibiotic-eluting, resorbable bone graft device with controlled, localized delivery of therapeutically relevant antibiotics that also supports new bone formation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Antibiotic-impregnated bone graft to prevent infection after total hip arthroplasty (ABOGRAFT): protocol for a randomised, double-blind, placebo-controlled trial.BMJ open · 2026
- Antibiotic-Loaded Ceramic Bone Grafts Exhibit Burst, Not Sustained, Gentamicin Release: In Vivo Evidence and Reanalysis of in Vitro Assay Artifacts Informing Clinical Use.The Journal of bone and joint surgery. American volume · 2026
- Impact of surface structuring on antibiotic elution from bone cement in two-stage revision for periprosthetic joint infection.BMC musculoskeletal disorders · 2026
- Double-Layered Antibiotic-Loaded Cement Spacer for Hip Infection: A Report of Two Cases.Journal of orthopaedic case reports · 2026
- Efficacy of Antibiotic-Loaded Stimulan Beads Against Biofilms on Clinically Relevant Orthopedic Implant Surfaces.Antibiotics (Basel, Switzerland) · 2026
- Aptamer-Targeted PLGA Nanoparticles for Biofilm-Directed Intra-articular Antibiotic Delivery in Prosthetic Joint InfectionbioRxiv · 2026