International journal of mass spectrometry

Fabrication of thermoplastic microfluidic probes for nanospray desorption electrospray ionization mass spectrometry.

Renkai Dong, Anthony H Choi, Xiangtang Li, Xindi Tang, Li-Xue Jiang, Julia Laskin

James Tarpo Jr. and Margaret Tarpo Department of Chemistry, Purdue University, West Lafayette, IN, 47907, United States.

Key points

From the paper's conclusion
  • Here, we present a mold-based fabrication protocol for cyclic olefin copolymer (COC) microfluidic probes (MFPs).

Abstract

Nanospray desorption electrospray ionization mass spectrometry (nano-DESI MS) uses specially designed probes to achieve a stable liquid bridge on the sample surface, ensuring consistent analyte extraction. Here, we present a mold-based fabrication protocol for cyclic olefin copolymer (COC) microfluidic probes (MFPs). We describe mold design, probe fabrication, and probe setup for nano-DESI MS. Four mold fabrication methods, including computer numerical control (CNC) machining, metal 3D printing, photolithography, and selective laser-assisted etching (SLE), were evaluated, with SLE providing the highest structural precision and enabling reproducible channel fabrication. Performance evaluation using lysophosphatidylethanolamine (LPE) 17:1 standard demonstrated stable liquid bridge formation, clear spatial separation, reproducible signal response, and low background interference. This mold-based fabrication strategy produces reproducible, low-cost probes and offers an accessible approach for nano-DESI MS.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers