Local Conformal Auditing Exposes and Restores Reliability of AI-ECG Atrial Fibrillation Detection Under External Deployment Shift.

Talha Abbasi, Ali Abbas, Umair Hamid Mehmood

Shien-Ming Wu School of Intelligent Engineering, South China University of Technology, Guangzhou, China. Electronic address: 202422800072@mail.scut.edu.cn. · School of Materials Science and Engineering, South China University of Technology, Guangzhou, China.

Key points

From the paper's conclusion
  • AI-ECG uncertainty can fail silently despite high AUC. A local conformal audit detects when uncertainty estimates are untrustworthy and recalibrates the uncertainty layer without changing the classifier, converting trust into a measurable, subgroup-aware, task-specific guarantee.

Abstract

BACKGROUND: AI-enabled electrocardiography (AI-ECG) can detect atrial fibrillation (AF), but safe deployment depends on calibration, subgroup reliability, and explicit error guarantees, not on discrimination alone. OBJECTIVE: To test whether the conformal-prediction coverage guarantee of an AF ECG classifier survives external deployment shift, and whether a small local labeled audit restores it without retraining. METHODS: A four-class 12-lead rhythm classifier trained on PTB-XL (Germany; 7,831 ECGs, 7,164 patients) was evaluated unchanged on Chinese (Chapman-Shaoxing; n=2,289) and US (Georgia; n=1,990) cohorts. We measured discrimination, expected calibration error (ECE), 90%-target split-conformal coverage, subgroup coverage, label-free shift corrections, local recalibration across audit sizes, and an AF auto-clear policy with finite-sample risk control. RESULTS: AF AUC remained 0.998 (95% CI 0.997-0.999) in China but fell to 0.936 (0.919-0.952) in the US cohort. ECE rose from 0.051 to 0.197 and 0.237, and source-calibrated coverage fell from 0.901 to 0.758 and 0.722 (both P<0.001). Coverage was lower in Chinese men than women (0.707 vs 0.822; difference 0.115, 95% CI 0.081-0.148). Label-free corrections failed. Local recalibration restored mean coverage to 0.90 with 50 labeled ECGs but with wide variability (95% simulation interval 0.82-0.97), narrowing to 0.86-0.94 at 200 ECGs. Local AF risk control certified auto-clearing of 68% of held-out ECGs in China but abstained in the US cohort. CONCLUSION: AI-ECG uncertainty can fail silently despite high AUC. A local conformal audit detects when uncertainty estimates are untrustworthy and recalibrates the uncertainty layer without changing the classifier, converting trust into a measurable, subgroup-aware, task-specific guarantee.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Local Conformal Auditing Exposes and Restores Reliability of AI-ECG Atrial Fibrillation Detection Under External Deployment Shift — Science Skim