Durability and reconnection patterns at time of repeat ablation after first-time pulmonary vein isolation by monopolar focal pulsed field ablation.
Martin H Ruwald, Martin Haugdal, Christian Jøns, Rene Worck, Christina Alhede, Arne Johannessen, Uffe Gang, Niels Sandgaard,
Division of Electrophysiology, Department of Cardiology, Copenhagen Heart Center, Rigshospitalet, Copenhagen, Denmark.
Key points
From the paper's conclusion- Approximately half of patients demonstrated durable PVI at repeat procedures. Reconnection gaps clustered mainly in carina regions, highlighting the need for further studies evaluating chronic lesion durability after PFA.
Abstract
BACKGROUND: Reconnection patterns and durability after pulmonary vein isolation (PVI) or cavotricuspid isthmus block (CTIB) using monopolar focal pulsed field ablation (FPFA) at repeat ablation have not previously been evaluated. OBJECTIVE: This study aimed to characterize electrophysiological findings during repeat procedures in real-world patients with atrial fibrillation (AF). METHODS: Patients undergoing repeat ablation (n = 23) for symptomatic recurrent arrhythmias after first-time FPFA-guided PVI (n = 212) between October 2022 and April 2026 underwent 3-dimensional high-density mapping and ad hoc re-ablation using radiofrequency or FPFA. RESULTS: At index ablation, persistent AF was present in 13 of 23 (57%) patients, and CTIB was performed in 13 of 23 (57%) patients. Mean time to repeat procedure was 379 ± 282 days. Complete durable PVI was observed in 11 of 23 patients (48%; 95% confidence interval [CI], 27%-69%), corresponding to durable isolation in 74 of 94 veins (79%; 95% CI, 69%-87%). Regular atrial tachycardia was the predominant mechanism for recurrence, occurring in 17 of 23 patients (74%; 95% CI, 52%-90%). 17 conduction gaps were identified, predominantly clustered in carina regions (13 of 17 [76%]; P = .002). Reconnection occurred exclusively ipsilaterally, without significant differences between veins (17%-26%, P = .796). Durable CTIB was observed in 9 of 13 patients (69%; 95% CI, 39%-91%). After repeat ablation, 15 of 23 patients (65%; 95% CI, 43%-84%) remained free from atrial arrhythmia recurrence during 285 ± 204 days of follow-up. CONCLUSION: Approximately half of patients demonstrated durable PVI at repeat procedures. Reconnection gaps clustered mainly in carina regions, highlighting the need for further studies evaluating chronic lesion durability after PFA.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Detailed characterization of pulmonary vein reconnection and af triggers during redo ablation after single-shot pulsed-field ablationEuropean Heart Journal · 2026
- Pulmonary Vein Reconnection Burden and Extra-Pulmonary Vein Ablation During Second Catheter Ablation for Atrial Fibrillation in the Pre-Pulsed-Field Ablation Era: A Multicenter Registry.Journal of cardiovascular electrophysiology · 2026
- Lesion durability after pulmonary vein isolation and linear lesions in complex atrial tachycardias at repeat ablations following focal pulsed field ablationEuropean Heart Journal · 2026
- Pulsed-field ablation demonstrates superior pulmonary vein isolation durability at 90 days: Pooled analysis of protocol-mandated remapping dataHeart Rhythm · 2025
- Characteristics of Recurrent Atrial Arrhythmias and Lesion Assessment Late After Pulsed Field Ablation of Atrial Fibrillation.Circulation. Arrhythmia and electrophysiology · 2026
- Characteristics of pulmonary vein reconnection sites across three pulsed-field ablation systems: findings from repeat proceduresEuropean Heart Journal · 2026