Hearing research

Efficiency of a vibrating cochlear implant array for acoustic stimulation.

Mohammad Ghoncheh, Daniel Schurzig, Nils Prenzler, Hannes Maier

Department of Otolaryngology and Institute of Audioneurotechnology (VIANNA), Hannover Medical School, Hannover, Germany; Cluster of Excellence "Hearing4all", Hannover Medical School, Hannover, Germany. Electronic address: Ghoncheh.Mohammad@MH-Hannover.de. · Department of Otolaryngology and Institute of Audioneurotechnology (VIANNA), Hannover Medical School, Hannover, Germany; MED-EL research center, Hannover, Germany.

Key points

From the paper's conclusion
  • The findings indicate that a 1 µm vibration of a CI electrode can provide an average output of ∼110 eq. dB SPL, potentially compatible with the output required for EAS.

Abstract

INTRODUCTION: Patients with conductive or mixed hearing loss can benefit from a combination of acoustic or mechanical stimulation and a cochlear implant (CI). We experimentally investigated the feasibility of using the CI electrode array as an acousto-mechanical stimulator in human temporal bones (TBs). METHODS: First, the displacement magnitude of the stapes in response to the acoustic stimulation in the external auditory canal was measured in temporal bone specimens (N = 10). The scala vestibuli (SV) intracochlear pressure (PSV) in response to the acoustic stimulation was measured. Next, a CI (FLEX20, MED-EL, Austria) electrode was inserted into the scala tympani (ST) through the round window, and its distal end was attached to a commercial piezoelectric transducer using a custom-made adapter. Finally, the electrode displacement and PSV was measured in response to mechanical stimulation of the inserted CI electrode. The equivalent sound pressure level (eq. dB SPL) at 1 µm electrode displacement was determined to evaluate the achieved output level of the vibrating CI. RESULTS: The vibrating CI produced an average equivalent output level of 108 eq. dB SPL (LDisp) at low frequencies (125 - 500 Hz) for a 1 µm electrode displacement of the CI. Above 1 kHz, the output increased by ∼10 dB per octave, reaching 103-139 eq. dB SPL, with a variability of 8-15 dB (SD) across experiments and frequencies up to 10 kHz. CONCLUSION: The findings indicate that a 1 µm vibration of a CI electrode can provide an average output of ∼110 eq. dB SPL, potentially compatible with the output required for EAS.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Efficiency of a vibrating cochlear implant array for acoustic stimulation — Science Skim