Efficiency of a vibrating cochlear implant array for acoustic stimulation.
Mohammad Ghoncheh, Daniel Schurzig, Nils Prenzler, Hannes Maier
Department of Otolaryngology and Institute of Audioneurotechnology (VIANNA), Hannover Medical School, Hannover, Germany; Cluster of Excellence "Hearing4all", Hannover Medical School, Hannover, Germany. Electronic address: Ghoncheh.Mohammad@MH-Hannover.de. · Department of Otolaryngology and Institute of Audioneurotechnology (VIANNA), Hannover Medical School, Hannover, Germany; MED-EL research center, Hannover, Germany.
Key points
From the paper's conclusion- The findings indicate that a 1 µm vibration of a CI electrode can provide an average output of ∼110 eq. dB SPL, potentially compatible with the output required for EAS.
Abstract
INTRODUCTION: Patients with conductive or mixed hearing loss can benefit from a combination of acoustic or mechanical stimulation and a cochlear implant (CI). We experimentally investigated the feasibility of using the CI electrode array as an acousto-mechanical stimulator in human temporal bones (TBs). METHODS: First, the displacement magnitude of the stapes in response to the acoustic stimulation in the external auditory canal was measured in temporal bone specimens (N = 10). The scala vestibuli (SV) intracochlear pressure (PSV) in response to the acoustic stimulation was measured. Next, a CI (FLEX20, MED-EL, Austria) electrode was inserted into the scala tympani (ST) through the round window, and its distal end was attached to a commercial piezoelectric transducer using a custom-made adapter. Finally, the electrode displacement and PSV was measured in response to mechanical stimulation of the inserted CI electrode. The equivalent sound pressure level (eq. dB SPL) at 1 µm electrode displacement was determined to evaluate the achieved output level of the vibrating CI. RESULTS: The vibrating CI produced an average equivalent output level of 108 eq. dB SPL (LDisp) at low frequencies (125 - 500 Hz) for a 1 µm electrode displacement of the CI. Above 1 kHz, the output increased by ∼10 dB per octave, reaching 103-139 eq. dB SPL, with a variability of 8-15 dB (SD) across experiments and frequencies up to 10 kHz. CONCLUSION: The findings indicate that a 1 µm vibration of a CI electrode can provide an average output of ∼110 eq. dB SPL, potentially compatible with the output required for EAS.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Noninvasive indirect intracochlear pressure monitoring during cochlear implantationHearing Research · 2026
- Acoustic Masking of Electric Stimulation at Basal and Extra-Cochlear SitesbioRxiv · 2026
- Influence of Cochlear Microanatomy on Hearing Preservation in Cochlear Implant Recipients.Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026
- Relationships between ECAP amplitude growth functions and perceptual growth in cochlear implant users: Implications for individualized cochlear implant fitting.Hearing research · 2026
- Novel psychoacoustic measure of spectral-temporal processing for discrete regions of the cochlear array: a proof-of-concept study.Audiology & neuro-otology · 2026
- Residual Hearing Loss After Chronic Cochlear Implantation in Guinea Pigs is Primarily Conductive.bioRxiv : the preprint server for biology · 2026