At-Home Critical Flicker Frequency Testing in Cirrhosis: A Feasibility Pilot Study Using the Novel Beacon Device.
Philip Vutien, Richard Li, Ravi Karkar, James Fogarty, Sean A Munson, Sabrina Omer, Michael Yacoub, James Kashima,
Division of Gastroenterology and Hepatology, University of Washington, Seattle, Washington. · Paul G. Allen School of Computer Science and Engineering, University of Washington, Seattle, Washington.
Key points
From the paper's conclusion- Beacon enabled reliable, self-administered CFF testing at home with high adherence and no evidence of learning effects. Further studies incorporating other validated MHE tools are needed to determine whether at-home CFF monitoring can detect MHE early and prevent overt HE episodes and hospitalizations.
Abstract
BACKGROUND AND AIMS: Critical flicker frequency (CFF) is a well-validated neurophysiologic screening test for minimal hepatic encephalopathy (MHE) in cirrhosis. We evaluated the feasibility and reliability of repeated at-home, self-administered CFF testing using a novel device (Beacon). METHODS: We enrolled 21 participants with cirrhosis. Each received a Beacon device and smartphone preloaded with an application to measure CFF using 2 protocols: (1) method of limits-descending (MOL-D) and (2) forced choice. Participants were instructed to complete ≥5 tests per week for 2 weeks, followed by ≥ 1 test per week for an additional 4 weeks. Outcomes included adherence, test duration, and longitudinal CFF trends to assess potential learning effects from repeated testing. RESULTS: Participants were predominantly male (76%) with a mean age of 52.2 years; 43% had alcohol-related liver disease and 38% had prior overt HE. Adherence (completing ≥70% of recommended measurements) was observed in 19 (90.5%) for daily and 16 participants (76.2%) for weekly testing. Median CFF was 42.9 Hz (interquartile range: 38.4-44.4) by MOL-D and 43.7 Hz (interquartile range: 41.7-45.8) by forced choice. Forced choice CFF was significantly lower in those with prior HE (median 40.7 Hz vs 44.9 Hz; P = .02), while MOL-D CFF did not differ by HE status (42.7 Hz vs 42.9 Hz; P = .44). CFF remained stable over time for both protocols (MOL-D: +0.01 Hz/day of testing, P = .67; forced choice: +0.03 Hz/day, P = .09). Forced choice testing required more time than MOL-D to complete (median 4.67 vs 2.48 minutes; P < .001). CONCLUSION: Beacon enabled reliable, self-administered CFF testing at home with high adherence and no evidence of learning effects. Further studies incorporating other validated MHE tools are needed to determine whether at-home CFF monitoring can detect MHE early and prevent overt HE episodes and hospitalizations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A novel 1-minute clamping bean test (CBT1) for screening covert hepatic encephalopathy: feasibility and diagnostic performance.Digestive and liver disease : official journal of the Italian Society of Gastroenterology and the Italian Association for the Study of the Liver · 2026
- Eye movement analysis detects early neurological alterations in cirrhotic patients classified as not having minimal hepatic encephalopathy by the Psychometric Hepatic Encephalopathy Score.PloS one · 2026
- Critical flicker fusion frequency assessment: robustness across photopic illumination in experienced divers.Frontiers in physiology · 2026
- Prediction of incident overt hepatic encephalopathy using individual and composite measures of minimal hepatic encephalopathy: The PrO-MHE Study.Annals of hepatology · 2026
- Development and clinical evaluation of the modified Hepatic Encephalopathy Staging Tool.Clinical gastroenterology and hepatology : the official clinical practice journal of the American Gastroenterological Association · 2026
- Combined multimodal monitoring during neonatal transition in preterm infants: feasibility, safety and post-implementation evaluation.European journal of pediatrics · 2026