Fabrication of ultrasensitive ciprofloxacin electrochemical aptasensor based on AuNPs decorated MWCNTs@GO@CS film.

Xiao-Zhen Feng, Yu-Qi Ge, Jiang-Tao Xu, Zhencheng Chen, Guo-Cheng Han, Zhiliang Jiang

School of Life and Environmental Sciences, Guangxi Human Physiological Information Non Invasive Detection Engineering Technology Research Center, Guangxi Key Laboratory of Metabolic Reprogramming and Intelligent Medical Engineering for Chronic Diseases, Guangxi Colleges and Universities Key Laboratory of Biomedical Sensors and Intelligent Instruments, Guilin University of Electronic Technology, Guilin, 541004, PR China. Electronic address: fxz97118@guet.edu.cn. · School of Life and Environmental Sciences, Guangxi Human Physiological Information Non Invasive Detection Engineering Technology Research Center, Guangxi Key Laboratory of Metabolic Reprogramming and Intelligent Medical Engineering for Chronic Diseases, Guangxi Colleges and Universities Key Laboratory of Biomedical Sensors and Intelligent Instruments, Guilin University of Electronic Technology, Guilin, 541004, PR China.

Key points

From the paper's conclusion
  • Moreover, results from chicken samples analyzed by the proposed sensor were consistent with HPLC-MS, with only 0.9% relative error, confirming the sensor's reliability and practical applicability for real-world CIP monitoring.

Abstract

A highly sensitive electrochemical aptasensor (BSA/Apt/MWCNTs@GO@CS/. AuNPs/SPCE) was fabricated for the efficient detection of ciprofloxacin (CIP). The complementary structural and physicochemical properties of MWCNTs and GO provide a conductive and functionalized platform for aptamer immobilization, while the excellent film-forming property of CS ensures stable and uniform deposition of the nanomaterials on the limited SPCE surface. The sensor displayed excellent linearity by the DPV method from 0.03 to 100.00 nM with a detection limit of 0.01 nmol/L. Spiked recovery tests in lake water and river water achieved 97.98-103.98% recovery with relative standard deviations below 10%, indicating acceptable precision. Moreover, results from chicken samples analyzed by the proposed sensor were consistent with HPLC-MS, with only 0.9% relative error, confirming the sensor's reliability and practical applicability for real-world CIP monitoring.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers