Experimental eye research

Ring-Based OCT Protocol for Retinal Ganglion Cell Complex Thickness Analysis in Mice.

Gil Ben-David, Shmuel Kagasov, Mohammad Darawashe, Moamen Masalha, Yael Nisgav, Ivan Budnik, Tami Livnat, Alon Zahavi

Ophthalmology Department and Eye Research Laboratory, Felsenstein Medical Research Center, Rabin Medical Center, Petach Tikva, Israel. · Ophthalmology Department and Eye Research Laboratory, Felsenstein Medical Research Center, Rabin Medical Center, Petach Tikva, Israel; Gray Faculty of Medical and Health Sciences, Tel Aviv University, Tel Aviv, Israel.

Preclinical

Key points

From the paper's conclusion
  • This protocol provides a structured approach for OCT-based GCC thickness assessment in mice and may improve consistency and comparability across preclinical studies.

Abstract

Optical coherence tomography (OCT) enables non-invasive, high-resolution imaging of retinal layers and is widely used to assess structural changes in preclinical models of retinal and optic nerve disease. However, OCT-based retinal thickness measurements in rodents remain insufficiently standardized, with differences in measurement location and segmentation methodology potentially limiting reproducibility and comparability across studies. We describe a standardized OCT-based protocol for ganglion cell complex (GCC) thickness measurement in mice using predefined retinal landmarks and manual segmentation. Six radial OCT scans separated by 30° are acquired around the optic nerve head, and GCC thickness is measured using ImageJ at three nominal radial distances on each side of the optic nerve: 1000, 1500, and 2000 μm. The applicability of the protocol was evaluated in a murine optic nerve crush (ONC) model. GCC thickness decreased significantly over time, from 68.8 ± 5.2 μm at baseline to 60.7 ± 8.0 μm at day 15 and 60.6 ± 9.8 μm at day 24. Radius- and quadrant-specific analyses further demonstrated the ability of the protocol to detect regional patterns of GCC thinning. This protocol provides a structured approach for OCT-based GCC thickness assessment in mice and may improve consistency and comparability across preclinical studies.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Ring-Based OCT Protocol for Retinal Ganglion Cell Complex Thickness Analysis in Mice — Science Skim