European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery

A New Dedicated Platform in the Treatment of Iliac and Femoropopliteal Lesions via Transradial Access.

Raphaël Coscas, Martin Werner, Paolo Sbarzaglia, Klaus A Hausegger, Gilles Goyault, Marianne Brodmann, Mercedes Guerra, Oliver Schlager,

Department of Vascular Surgery, Ambroise Paré University Hospital, Assistance Publique-Hôpitaux de Paris, Boulogne-Billancourt, France; UMR 1018, Inserm-Paris11 - CESP, Versailles Saint-Quentin-en-Yvelines University, Paris-Saclay University, Paul Brousse Hospital, Villejuif, France; Department of Vascular Surgery, American Hospital of Paris, Neuilly-sur-Seine, France. Electronic address: rcoscas@gmail.com. · Hanusch Krankenhaus, Vienna, Austria.

Key points

From the paper's conclusion
  • This new dedicated platform for transradial access in lower limb interventions met its primary safety and efficacy endpoints in this first clinical evaluation.

Abstract

OBJECTIVE: The results of a clinical study with a new dedicated platform comprising a guiding sheath, balloon catheter, and stent delivery system designed for transradial access for lower limb transluminal angioplasty are reported. METHODS: The RADIANCY trial is a European, multicentre, prospective, single arm study evaluating the RADIANZ Vascular Stent System (Cordis) in patients with symptomatic iliac or femoropopliteal disease. The primary safety endpoint was major radial access complications through discharge. The primary efficacy endpoint was technical success of the index procedure. Using the Clopper-Pearson method, a cohort of 129 patients provided 88% power to show technical success > 93% and complication rate < 7%. After a roll in phase, safety and efficacy were assessed in the main cohort. Adverse events were adjudicated by an independent clinical events committee. RESULTS: Across 12 European centres, 151 patients were enrolled, including 22 roll in patients and 129 in the main cohort. Patients were aged 68.2 ± 9.4 years, 26.4% were women, and their Rutherford classification score was 2.78 ± 0.74. Multiple vessel segments were treated, mainly in the common iliac and superficial femoral arteries. Through discharge, there were no major radial access site complications (0%, 95% confidence interval [CI] 2.30%), and the technical success in patients receiving the study stent system was 99.2% (95% CI 96.20%). Thus, the primary safety and efficacy endpoints were achieved. The thirty day peri-procedural adverse event rate was 18.6% (24 of 129), with transradial access site haematoma (n = 10) being the most common. Four (3.1%) serious device or procedure related adverse events through thirty days were reported, including one ischaemic stroke that recovered (0.8%), one superficial femoral artery dissection (0.8%) ultimately requiring index limb amputation (0.8%), and one grade II haematoma (0.8%). Patient discharge within twenty-four hours was 76.0% (98 of 129). CONCLUSION: This new dedicated platform for transradial access in lower limb interventions met its primary safety and efficacy endpoints in this first clinical evaluation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A New Dedicated Platform in the Treatment of Iliac and Femoropopliteal Lesions via Transradial Access — Science Skim