Fully automated, fast, global noise level calculation for abdomino-pelvic adult CT images.
Eric Pace, Carmel J Caruana, Hilde Bosmans, Kelvin Cortis, Melvin D'Anastasi, Gianluca Valentino
Medical Physics, Faculty of Health Science, University of Malta, Msida MSD2080, Malta. Electronic address: eric.pace@um.edu.mt. · Medical Physics, Faculty of Health Science, University of Malta, Msida MSD2080, Malta.
Key points
From the paper's conclusion- A free, fully-automated flexible tool for GNL calculation for adult abdomino-pelvic CT was developed. Since GNL values are dependent on choice of input parameters, these should therefore be reported alongside GNL values. Recommended parameters are threshold fences of - 300:0HU for adipose and 0:170HU for soft tissue (higher when contrast is present), 3 mm radius disc-kernel, 1.0HU bin-width.
Abstract
PURPOSE: The Global Noise Level (GNL) is an established objective CT noise metric evaluated directly on patient clinical images. A free, open-source, customisable tool to acceptance-test and validate commercial GNL software applications, monitor patient image noise, optimise protocols and which permits flexible input parameters for research purposes is lacking. This working developed such a tool and investigated the impact of various user-supplied input parameters on the GNL. METHOD: The tool was developed in Python and followed standard published methods; it was validated against manual calculation. The impact of user-supplied input parameters on GNL values, namely the upper/lower CT-number threshold-fences, kernel size and shape (square and novel disc-shape) and histogram bin-widths, was studied by applying the tool to two abdomino-pelvic multi-centric, multi-system and multi-phasic image datasets (n = 357). RESULTS: The tool accurately calculated GNL (Bland-Altman plot of automatic vs manual calculation showed bias = 0 and standard deviation = 0) with a processing rate of 14 slices/s. GNL values changed significantly as a function of threshold-fence. GNL values increased monotonically with kernel size for both square and disc kernels with the latter consistently giving lower values. GNL values also increased with bin-width. CONCLUSIONS: A free, fully-automated flexible tool for GNL calculation for adult abdomino-pelvic CT was developed. Since GNL values are dependent on choice of input parameters, these should therefore be reported alongside GNL values. Recommended parameters are threshold fences of - 300:0HU for adipose and 0:170HU for soft tissue (higher when contrast is present), 3 mm radius disc-kernel, 1.0HU bin-width.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- AnatomIQ: An Open-Source Toolkit for Automated Background Detection in Medical ImagingarXiv · 2026
- Automated patient-derived noise power spectrum estimation in abdominal CT: a preliminary validationBiomedical Physics & Engineering Express · 2026
- The Accuracy of Noise Measurements in Patient CT Images.Academic radiology · 2026
- Development of an Electronic Clinical Quality Measure for Excessive Radiation Dose or Inadequate Image Quality in Diagnostic CT in Adults.Journal of the American College of Radiology : JACR · 2026
- Automated measurement of detectability index in CT imaging: Development and validationJournal of Applied Clinical Medical Physics · 2025
- Shapley effects to identify the most influential input parameters of organ dose estimation following CT scans in childhood.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026