Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)

Efficient proton adaptive breast radiotherapy with CBCT-Based synthetic CT.

Emma Viviers, Julie A Bradley, Christina Klassen, Raymond Mailhot Vega, Nancy Mendenhall, Mark Artz, Hardev Grewal, Perry Johnson,

University of Florida Health Proton Therapy Institute, Jacksonville, FL, USA; Mayo Clinic, Jacksonville, FL, USA. · University of Florida Health Proton Therapy Institute, Jacksonville, FL, USA; Department of Radiation Oncology, University of Florida College of Medicine, Gainesville, FL, USA.

Observational · n=25

Key points

From the paper's conclusion
  • Accurate proton-dose calculations require reCT, but synthetic CT can effectively flag tissue alterations. This study shows that monitoring dosimetric shifts from vCT can prompt targeted reCT-based replanning in unilateral breast proton therapy, thereby sparing patients the exposure and logistics of routine reCT during treatment.

Abstract

PURPOSE: Repeat CTs (reCTs) are routinely used during breast proton radiotherapy to assess anatomical and positioning changes; however, they increase patient dose and workflow burden. This study evaluated CBCT-based synthetic CT imaging for accurately monitoring fractional dose distribution and developed an adaptive workflow for breast proton therapy. METHODS: A retrospective analysis included 26 breast cancer patients who received both CBCT and reCT on the same day between 2022 and 2024. Two commercial synthetic CT algorithms were studied: corrected CBCT (corrCBCT), which utilizes iterative correction, and virtual CT (vCT), generated by deforming the planning CT to match the CBCT. Treatment plans were recalculated on the corrCBCT and vCT, and compared with same-day reCT metrics. A prospective evaluation of weekly vCT-guided dosimetry was performed in an additional 25 patients to assess its feasibility in adaptive workflows. RESULTS: CorrCBCT and vCT produced reliable dosimetric results when compared with reCT. Target coverage, measured by the mean percentage dose difference in D95% to the breast or chest wall, was -0.80% (SD = 1.3) for the corrCBCT and -0.7% (SD = 1.6) for the vCT. vCT also showed good agreement with same-day reCT imaging for dose metrics of adjacent critical structures. Larger discrepancies were observed in nodal targets, primarily due to CBCT image artifacts in the shoulder region. CONCLUSION: Accurate proton-dose calculations require reCT, but synthetic CT can effectively flag tissue alterations. This study shows that monitoring dosimetric shifts from vCT can prompt targeted reCT-based replanning in unilateral breast proton therapy, thereby sparing patients the exposure and logistics of routine reCT during treatment.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers