Single prior-based 3D volumetric tracking from a planar snapshot in spine SBRT: A proof-of-concept study.
Euidam Kim, Donghyeok Choi, Yoonsun Chung, Hugh Lee
Department of Radiation Oncology, Washington University in St Louis School of Medicine, St Louis, MO, USA; Department of Nuclear Engineering, Hanyang University, Seoul, Republic of Korea; Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, USA. · Department of Radiation Oncology, Washington University in St Louis School of Medicine, St Louis, MO, USA.
Key points
From the paper's conclusion- This proof-of-concept study establishes the technical feasibility of data-efficient, patient-specific onboard volumetric tracking using a single volumetric prior by successfully resolving anatomical deformations from a single 2D projection in near real-time.
Abstract
BACKGROUND: Continuous 3D onboard monitoring is essential to ensure accurate dose delivery of stereotactic body radiation therapy. However, current approaches for 2D-to-3D volumetric estimation typically require extensive patient-specific data, or depend on auxiliary imaging hardware. This study introduces a proof-of-concept framework for data-efficient, patient-specific 3D volumetric tracking using only a single volumetric prior. METHODS: A retrospective dataset of 80 spine SBRT patients was utilized. Digitally Reconstructed Radiographs synthesized from cone-beam computed tomography volumes served as surrogates for onboard kV projections. The pipeline consists of patient-agnostic pre-training using an archival dataset, patient-specific adaptation using the single volumetric prior supplemented by non-rigid motion simulations, and simulated intra-fractional 3D estimation. The network employs a modified Attention U-Net with a channel-to-depth bottleneck and a triple feature concatenation to map the 2D to 3D spatial domain. RESULTS: The proposed model demonstrated a statistically significant improvement in structural tracking capabilities compared to a static uncompensated baseline. Furthermore, the performance variance across patients was substantially compressed, enhancing reliability against setup uncertainties while achieving high spatial concordance in the complex spinal anatomy. The inference throughput reached approximately 10 frames per second. CONCLUSIONS: This proof-of-concept study establishes the technical feasibility of data-efficient, patient-specific onboard volumetric tracking using a single volumetric prior by successfully resolving anatomical deformations from a single 2D projection in near real-time.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Just-in-time volumetric imaging: Learning 2D-3D thoracic motion models during image-guided interventionsarXiv · 2026
- Rapid patient-specific neural networks for X-ray to volume registration.Nature · 2026
- Assessing the impact of knowledge-based planning on dosimetric performance in adaptive prostate stereotactic body radiation therapy.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- Real‐time intrafraction 3D spine tracking using gantry‐mounted kV imaging for single fraction high dose functional radiosurgeryMedical Physics · 2026
- Anticipatory Digital Twins for Online Head-and-Neck Adaptive Proton Therapy via Foundation-Model Registration.ArXiv · 2026
- Streamlining Liver SABR: Validation and Initial Experience of Bulk Density Override Based Single-Session, Simulation-Free Adaptive Workflow.Practical radiation oncology · 2026